Labo Cube External Sealed System Cooling Water Circulator |
CF320S/320P |
|
|
External sealed system cooling water circulator with abuilt-insolvent collection device that can be installed under adraftchamber or a sink and is compatible withLaboCube. |
CF320S / CF320P | CF320S hasabuilt-insolvent collection device while CF320P has asolventcollectiondevice that has an external link mechanism to stopthecirculationpump through the connection to the Teflondiaphragmvacuum pump andan external device. |
|
Specifications |
Product code | 221551 | 221553 | |
Model | CF320S | CF320P | |
System | Externalsealed circulation | ||
Performance*1 | |||
Operating temp. range | -20deg.Cto room temp. | ||
Temp.adjustment accuracy | +/-2deg.C | ||
Cooling capacity | Approx.450W (385 Kcal/h) at liquid temp: 10 deg.C Approx. 350W (300 Kcal/h) at liquid temp: 0 deg.C Approx. 260W (220 Kcal/h) at liquid temp: -10 deg.C | ||
Circulationdeviceconfiguration | |||
Temperature control | Refrigerator ON-OFF control | ||
Temperature sensor | T-thermocouple (with SUS protective tube) | ||
Tempand time setting | Digitalsetting and LED display | ||
Refrigerator/refrigerant | Air-cooled 350W / R404A | ||
Circulation pump | Withamagnet pump and an H bypass | ||
Pump performance | Max. flow (50/60 Hz) | 15L/min | |
Max. lift (50/60 Hz) | 5.5/7.0m | ||
Unit circulation performance | Max. flow (50/60 Hz) | 9.1/8.9L/min | |
Max. lift (50/60 Hz) | 4.8/5.5m | ||
Cooling coil | Coppernickel plating | ||
Leg | Casterwheels | ||
Collectiondeviceconfiguration | |||
External circulation nozzle | O.D.10.5mm (SUS) I/O one system, with a discharge sidestopvalve | ||
Operating environmental temp. range | 5to35deg.C | ||
Solvent collection cooloing pipe | SUS304 Dia.8 Surface area: 530 cm2 | ||
Solvent collection nozzle | O.D.8mm(SUS) Each of I/O x 1 | ||
Solvent collecting bottle | 500mLglass bottle x 1 | ||
Connection to vacuum pump | - | ATeflondiaphragmvacuum pump PG200 with displacement of 20 L/min canbeconnected tothe inside of the main unit, with adedicatedserviceoutlet. | |
Function | |||
External link function | - | Externallinkterminal(no-voltage type): Circulation pump of the circulator is OFF and vacuum pumpisOFFwhen the contact is closed. | |
- | REsignalconnector: Vacuum pump delayed stop circuit with linked operationofRE600/800(vacuum pump delay timer communication lineisoptional) | ||
Otherfunctions | Drain,overflow | ||
Safety functions | Overcurrent electrical leakage breaker, refrigerator overloadrelay,pump thermal protector delay timer forprotectingrefrigerator, Hbypass for protecting pump | ||
Standards | |||
Watertank | Stainless steel, rectangle type, Approx. 4L (liquidamount:3L)With over flow | ||
Powersource | AC100V,10A (service outlet 5A) 50/60 Hz | ||
External dimensions*2 | 400x400x 600 mm (W/D/H) | ||
Weight | Approx.40 kg (excluding vacuum pump) | ||
Accessory | |||
Accessories | Topcover, drain hose, operation manual,warrantycard |
*1Conditions: temperature and humidity: 23deg.C+/-5deg.C,65%RH+/-20% (no load) *2 External dimensions donotincludeprotrusions. |
Installation example |
|
Circulation heat insulation hose (optional) | Vacuum hose (optional) | Vacuum pump PG200 (includes CF320P as standard) |
Displacement: 20/24 L/min (50/60 Hz) |
Solvent collection bottle | Cooling curve CF320S/320P | Flow/lift curve CF320S/320P |
Example system configuration (connection example between CF320P and RE600/800) |
Configuration of CF320P | Optional items | Dimensional drawing (mm) | |||||||||||||
|
|