pl2250皮秒激光器-激光器件

pl2250皮秒激光器-激光器件

参考价: 面议

具体成交价以合同协议为准
2020-09-04 01:53:19
191
产品属性
关闭
上海尖丰光电技术有限公司

上海尖丰光电技术有限公司

免费会员
收藏

组合推荐相似产品

产品简介

型号:PL2250

详细介绍

PL2250系列高能量皮秒锁模Nd:YAG激光器(PL2250 series high energy picosecond mode-locked Nd:YAG lasers)

品牌:EKSPLA

型号:PL2250

采用了单脉冲能量100 mJ,小于30皮秒(20 ps可选)脉冲持续时间,pl2250系列皮秒锁模钕:YAG激光器在高脉冲能量皮秒激光器建立了新标准。创新和成本效益的设计,可以提高激光的可靠性,同时降低运行和维护成本。

Features

  • Hermetically sealed DPSS master oscillator
  • Diode pumped regenerative amplifier
  • Flashlamp pumped power amplifier producing up to 100 mJ per pulse at 1064 nm
  • 30 ps pulse duration (20 ps optional)
  • Excellent pulse duration stability
  • Up to 50 Hz repetition rate
  • Streak camera triggering pulse with <10 ps rms jitter
  • Excellent beam pointing stability
  • Thermo stabilized second, third or fourth harmonic generator options
  • PC control via USB and LabVIEW? drivers (RS232 control is optional)
  • Remote control via keypad

Applications

  • Time resolved spectroscopy
  • SFG/SHG spectroscopy
  • Nonlinear spectroscopy
  • OPG pumping
  • Remote laser sensing
  • Satellite ranging
  • Other spectroscopic and nonlinear optics experiments

PL2250 series lasers set a new standard in high pulse energy picosecond lasers. Their innovative and cost-effective design improves laser reliability and reduces running and maintenance costs.

Innovative design

The heart of the system is a diode pumped solid state (DPSS) master oscillator placed in a hermetically sealed monolithic block. The flashlamp pumped regenerative amplifier is replaced by an innovative diode pumped regenerative amplifier. Diode pumping results in negligible thermal lensing, which allows operation of the regenerative amplifier at variable repetition rates, as well as improved long-term stability and maintenance-free operation.
The optimized multiple-pass power amplifier is flashlamp pumped and is optimized for efficient amplification of pulse while maintaining a near Gaussian beam profile and low wavefront distortion. The output pulse energy can be adjusted in approximately 1% steps, at the same time as pulse-to-pulse energy stability remains less than 0.8% rms at 1064 nm. Angle-tuned KD*P and KDP crystals mounted in thermostabilised ovens are used for second, third and fourth harmonic generation. Harmonics separators ensure the high spectral purity of each harmonic directed to different output ports.
Built-in energy monitors continuously monitor output pulse energy. Data from the energy monitor can be seen on the remote keypad or PC monitor. The laser provides several triggering pulses for synchronization of the customer‘s equipment. The lead or delay of the triggering pulse can be adjusted in 0.25 ns steps from the control pad or PC. Up to 1000 μs lead of triggering pulse is available as a pretrigger option.
Precise pulse energy control, excellent short-term and long-term stability, and up to 50 Hz repetition rate makes PL2250 series lasers an excellent choice for many demanding scientific applications.

Simple and convenient laser control

For customer convenience the laser can be controlled from a user-friendly remote control pad or USB interface. The remote pad allows easy control of all parameters and features a backlit display that is easy to read even while wearing laser safety eyewear. Alternatively, the laser can be controlled from a personal computer with supplied software for a Windows? operating system. LabVIEW? drivers are supplied as well.

ModelPL2250PL2251PL2251APL2251BPL2251C
Pulse energy
At 1064 nm 1 mJ 30 mJ 50 mJ 2) 80 mJ 2) 100 mJ
At 532 nm 3) 0.45 mJ 15 mJ 25 mJ 40 mJ 50 mJ
At 355 nm 4) 0.3 mJ 10 mJ 15 mJ 24 mJ 30 mJ
At 266 nm 5) 0.15 mJ 3 mJ 7 mJ 10 mJ 12 mJ
At 213 nm 6)-1.5 mJ3 mJ4.5 mJ6 mJ
Pulse energy stability, (Std. Dev.) 7)
At 1064 nm < 0.2 % < 0.8 %
At 532 nm < 0.4 % <1.0 %
At 355 nm < 0.5 % < 1.1 %
At 266 nm < 0.5 % < 1.2 %
At 213 nm- < 2.5 %
Pulse duration (FWHM) 8) 30 ± 3 ps
Pulse duration stability 9) ± 1.0 ps
Repetition rate 0-50 Hz50 or 10 Hz50, 20 or 10 Hz20 or 10 Hz10 Hz
Polarizationlinear, vertical, > 100:1
Pre-pulse contrast > 200:1
Triggering mode internal/external
SYNC OUT pulse jitter 10) < 0.1 ns
SYNC OUT pulse delay 11) - 500...50 ns
Beam divergence 12) < 1.5 mrad< 0.6 mrad< 0.5 mrad< 0.5 mrad< 0.5 mrad
Beam pointing stability 13) < 10 μrad< 50 μrad< 50 μrad< 50 μrad< 50 μrad
Beam diameter 14) ~2.5 mm~ 6 mm~ 8 mm~10 mm~12 mm
Typical warm-up time5 min30 min30 min30 min30 min

Physical characteristics

ModelPL2250PL2251PL2251APL2251BPL2251C
Laser head size (W x L x H)453x1024x260 mm453x1224x250 mm
Electric cabinet size (W x L x H)12 VDC power adapter, 85x170x41 mm550x600x550 mm
Umbilical length2.5 m

Operating requirements

ModelPL2250PL2251PL2251APL2251BPL2251C
Water consumption (max 20 °C )Air cooled< 8 l/min
Room temperature22 ± 2 °C
Relative humidity20-80 % (non-condensing)
Power requirements 15)110-240 VAC, 50/60 HzSingle phase, 208 or 230 VAC, 16A, 50/60 Hz
Power 16)< 0.15 kVA< 1.5 kVA< 1.5 kVA< 2 kVA< 2.5 kVA

1) Due to continuous improvement, all specifications are subject to change without notice. Parameters marked typical are not specifications. They are indications of typical performance and will vary with each unit we manufacture. Unless stated otherwise, all specifications are measured at 1064 nm.

2) PL2251A-50 has 40 mJ at 1064 nm, PL2251B-20 has 70 mJ at 1064 nm output energy. Inquire for these energies at other wavelengths.

3) For -SH option. Outputs are not simultaneous. Please inquire for pulse energies at other wavelengths.

4) For -TH option. Outputs are not simultaneous. Please inquire for pulse energies at other wavelengths.

5) For -FH option. Outputs are not simultaneous. Please inquire for pulse energies at other wavelengths.

6) For -FiH option. Outputs are not simultaneous. Please inquire for pulse energies at other wavelengths. FiH option is supplied in separate harmonics unit.

7) Averaged from 300 pulses.

8) Inquire for optional pulse durations in 20 – 90 ps range.

9) Measured over 1 hour period when ambient temperature variation is less than ±1 °C.

10) 10 ps jitter is provided with PRETRIG option.

11) SYNC OUT lead or delay can be adjusted with 0.25 ns steps in specified range.

12) Full angle measured at the 1/e2 point at 1064 nm .

13) RMS value measured from 300 shots.

14) Beam diameter is measured at 1064 nm at the 1/e2 point.

15) Three phase 208 or 380 VAC mains are required for 50 Hz versions.

16) For 10 Hz version.

上一篇:兄弟激光器不出光(维修)基本方案 下一篇:EO激光模块(维修)温度过高
热线电话 在线询价
提示

仪表网采购电话