品牌
生产厂家厂商性质
北京市所在地
ZJC-50KV介电强度测试机
一、用途
First, the use of
ZJC-50KV介电强度测试机采用计算机控制,通过人机对话方式,完成对绝缘介质的工频电压击穿,工频耐压试验。主要适用于固体绝缘材料如绝缘漆、树脂和胶、浸渍纤维制品、云母及其制品、塑料、薄膜复合制品、陶瓷和玻璃等介质在工频电压或直流电压下击穿强度和耐电压时间的测试;采用计算机控制,可对试验过程中的各种数据进行快速、准确的采集、处理,并可存取、显示、打印。
ZJC-50KV controlled by computer, through the man-machine dialogue mode, the voltage breakdown insulation, power frequency withstand voltage test. Mainly applicable to solid insulating materials such as insulation paint, resin and rubber, impregnated fiber products, mica products, plastic film, and composite products, ceramic and glass medium in the frequency of the voltage or DC voltage breakdown strength and voltage withstand the test of time; controlled by computer, the various test data in the process of rapid, accurate data acquisition, processing, and storage, display and print.
二、依据标准
Two, according to the standard
1、GB1408.1-2006《绝缘材料电气强度试验方法》
1, GB1408.1-2006 "electrical strength of insulating materials test method"
2、GB1408.2-2006《绝缘材料电气强度试验方法
2, "GB1408.2-2006 electrical strength of insulating materials test methods
第2部分:对应用直流电压试验的附加要求》
Part second: Additional requirements for tests using direct voltage "
3、GB/T1695-2005《硫化橡胶工频击穿电压强度和耐电压的测定方法》
3, GB/T1695-2005 "vulcanized rubber power frequency breakdown voltage strength and resistance to voltage measurement method"
4、ASM D149《固体电绝缘材料工业电源频率下的介电击穿电压和介电强度的试验方法》
4. D149 ASM test method for dielectric breakdown voltage and dielectric strength of solid electrical insulating material industrial power frequency
三、功能:
Three, function:
1、试验过程中可动态绘制出试验曲线,试验的曲线可以多种颜色叠加对比。
1, the test process can be dynamically drawn out of the test curve, the curve can be compared to a variety of color overlay.
2、可对试验数据进行编辑修改,灵活适用;
2, the test data can be edited and modified, flexible;
3、试验条件及测试结果等数据可自动存储;
3, test conditions and test results and other data can be stored automatically;
4、试验报告格式灵活可变,适用于不同用户的不同需求;
4, test report format flexible, applicable to different needs of different users;
5、可对一组试验中曲线数据的有效与否进行人为选定;
5, can be used for a group of test data in the curve of the effective or not artificially selected;
6、试验结果数据可导入EXECL,WORD文档编辑;
6, the results of the test data can be imported into WORD, EXECL document editing;
7、过电流保护装置有足够的灵敏度,能够保证试样击穿时在0.1S内切断电源;
7, over current protection device with sufficient sensitivity, to ensure that the sample breakdown in the 0.1S cut off the power supply;
8、仪器运行的持久性: 仪器可连续运行使用,不需为保护仪器而定期停机。
8, the persistence of the operation of the instrument: the instrument can be used continuously, without the need for the protection of equipment and regular downtime.
四、主要技术参数:
Four, the main technical parameters:
1、输入电压: AC 220V
1, input voltage: AC 220V
2、输出电压AC: 0~50KV;DC:0~50KV
2, the output voltage of AC: 0 ~ 50KV; DC:0 ~ 50KV
3、输出功率: 5KVA
3, output power: 5KVA
4、测量范围: 0KV~50KV
4, measurement range: 0KV ~ 50KV
5、电压测量误差:≤ 2%
5, voltage measurement error is less than 2%.
6、升压速率: 0.1kV/s ~10 kV/s
6, pressure rate: 0.1kV/s ~ 10 kV/s
7、耐压试验电压:10~50KV连续可调
7, voltage withstand test voltage: 10 ~ 50KV continuously adjustable
8、耐压时间: 0~4H(无电流导通情况下)
8, pressure time: 0 ~ 4H (no current situation)
9、电源: 220V±10%的单相交流电压和50Hz±1%的频率
9 power: 220V + 10% 50Hz + 1% single-phase AC voltage and frequency
10、电源电压稳定度外界电源电压波动≤10%
10, the power supply voltage stability of the external power supply voltage fluctuation is less than or equal to 10%
11、长×宽×高700mm×800mm×1650mm
11, length * width * high 700mm * 800mm * 1650mm
12、设备自重: 100kg
12, equipment weight: 100kg
13、运行环境温度:15 ~ 30℃,相对湿度:30%~85%能够稳定运行。
13, operating environment temperature: 15 ~ 30 degrees, relative humidity: 30% ~ 85% can be stable operation.
14、接地要求仪器需要单独接地,接地附合国家标准要求
14, grounding requirements instruments need to be grounded alone, and then meet the requirements of the national standard
15、接地电阻要求≤4Ω(用户实验室自行准备)
15, grounding resistance less than 4 ohm (user laboratory preparation)
四、主要配置:
Four, the main configuration:
1、主机 一台
1, the host a
2、试验台 一个
2 test a
3、油盒 一个
3, oil box a
4、试验电极 三个
4, the test electrode three
5、自动放电装置 一套
5, automatic discharge device
6、试验软件 一套
6, a set of test software
7、计算机 一台
7, a computer
8、喷墨打印机 一台
8, a inkjet printer
9、电源线数据线 一套
9, a set of power line data
五、安全说明:
Five, safety instructions:
1、要安装单独的保护地线,主要是减少试样击穿时对周围产生的较强的电磁干扰。也可避免控制计算机失控。
1, to install a separate protective ground, mainly to reduce the breakdown of the sample around the strong electromagnetic interference generated. Can also avoid the control of computer control.
2、直流试验放电保护功能:在设备做完直流试验时,当开启试验门时设备会自动保护,直至使用设备上的放电装置放电后保护会自动取消.(注:因为直流试验后不放电会危险到人身安全,不能直接拿取电极,起到提醒使用人员放电以免造成伤害)。
2, DC discharge protection function: test equipment in DC test finished, the device will automatically open the door when the protection test device, discharge device after discharge until the use of protection will be automatically canceled. (Note: because the DC test not put electricity danger to personal safety, can not directly take to remind the use of electrodes. Discharge of personnel so as to avoid injury).
3、增配试验手动放电装置,随主机为一体化,当直流试验过程中突然断电,可采用手动放电棒进行放电,保证试验人员的人身安全。
3, equipped with a manual test discharge device, with the host for the integration, when sudden power DC test process, can adopt manual discharging rod to discharge, ensure the personal safety of testing personnel.
4、该试验设备的电路设有多项保护措施,主要有:过流保护、失压保护、漏电保护、短路保护、直流试验放电保护等。
4, the circuit of the test equipment is equipped with a number of protective measures, mainly include: over current protection, voltage loss protection, leakage protection, short-circuit protection, DC test discharge protection, etc..
5、六级高压安全断电控制:
5, six level high voltage safety power control:
1、总电源开关
1, the total power switch
2、高压断电开关(钥匙开关)
2, high pressure switch (key switch)
3、调压器复位开关
3, voltage regulator reset switch
4、试验箱门安全开关
4 test box, door safety switch
5、高压变压器输入端限流空开
5, high voltage transformer input limit open flow
6、漏电保护开关
6, leakage protection switch