品牌
生产厂家厂商性质
上海市所在地
Application
This unit, mounted on an aluminium plate,
provides a compley pre-installed sample
gas conditioning for continuous use that can
be excellently integrated within gas analysis
systems.
Its compact construction only takes up little
space. The SS-5.. units are ready for use in a
few minutes. This makes time-consuming
procurement of individual components and
assembly superfluous.
Typical application examples for the SS-5..
units are: flue-gas and process-gas conditioning.
For special problems like aerosols, various solvents,
explosive gases in hazardous areas, we
can provide you with other solutions.
Description
The M&C SS-5... gas conditioning unit is
equiped with an ECP...... gas cooler which
cools the sample gas to constant +5 °C independend
of the ambient temperature.
As soon as the operating temperature of <
+8 °C is reached after start-up, the gas pump
N...KPE is switched on automatically via the
status contact of the gas cooler.
The SR 25.1 peristaltic pump ensures a constant
condensate removal, which makes a
long term measurement possible without
problems.
The corresponding particle filtration is carried
out by a 2μm filter type FP-2T.
This makes the SS-5... unit a complete gas
conditioning system suitable for most of the
analysers.