电加热接头T..-H1, V..-H1:
Application
In analysis technique, often temperatures
must be kept above the sample gas dew
point. Therefore, it is absoluy necessary to
avoid cold bridges.
In order to ensure this, the temperature regulated
M&C hose-/tube connectors T..-H1 and
V..-H1 are used to connect heated sample
lines type 3/4/5-N/M/H with an operating
temperature up to 180 °C.
Description
The M&C hose-/tube connectors T..-H1 and
V..-H1 are fixed on a mounting plate, decoupled
from heat and covered with an insulated
enclosure.
The heater consists of a heating element
with high capacity. The temperature is adjustable
on the integrated thermostat up to 180
°C with high temperature limiter and low
temperature alarm. The connection box with
integrated thermostat is installed outside the
enclosure on the mounting plate.
The bushings for the heated sample lines in
the enclosure are insulated with sockets of
heat resistant silicone. In order to avoid cold
bridges, the connector is compley heated
by means of a double-ended thermal conducting
jaw. For fixing the electrically heated
sample lines type 3/4/5-N/M/H – see data
sheet 3.1 –, mounting brackets are integrated.