Mcpherson Flat Field Spectrograph

Mcpherson Flat Field Spectrograph

参考价: 面议

具体成交价以合同协议为准
2023-03-08 08:33:44
92
产品属性
关闭
北京埃光科技有限公司

北京埃光科技有限公司

免费会员
收藏

组合推荐相似产品

产品简介

Aberration corrected toroidal gratingsEasy coupling of flat detectors (MCP or CCD)Flat field for direct detection with CCDsRapid spectral data collectionUltra High Vacuum (UHV)

详细介绍

商品描述

Aberration corrected toroidal gratings

Easy coupling of flat detectors (MCP or CCD)

Flat field for direct detection with CCDs

Rapid spectral data collection

Ultra High Vacuum (UHV)



· Aberration corrected toroidal gratings

· Easy coupling of flat detectors (MCP or CCD)

· Flat field for direct detection with CCDs

· Rapid spectral data collection

· Ultra High Vacuum (UHV)


Vacuum ultraviolet spectrograph collects a wide wavelength range in one acquisition. Three different diffraction gratings are available for work from about 10nm to 160nm. Each offers different spectral resolution and spectral range; they work in the soft x-ray (SXR, XUV), extreme ultraviolet (EUV), and VUV. This is a practical instrument for applications requiring rapid data collection from transient, short lived events. Examples include fusion plasmas formed in tokamak devices, laser ablation, liquid droplet EUVL sources, and Xenon plasma at 13.5nm considered for extreme ultraviolet lithography. 

The flat field toroidal grating instruments works over a fixed wavelength range. The prescription of the toroid and aberration correction determines the optimum region of use. Due to intensive design and difficult optical fabrication selection of gratings is limited. If there is a grating available for your wavelength of interest, you are in luck! These instruments exhibit relatively simple opto-mechanical systems that deliver good imaging and fair spectral resolution. They also allow use of a direct detection CCD to acquire high-energy spectra. Fragile microchannel plate (MCP) intensifiers can be reserved for those systems requiring gating.


Specifications

Optical DesignToroidal Grating Flat Field Spectrograph
Angle of Incidence~71 degrees
Focal Length292.1 mm
Acceptance24 mrad
Wavelength Rangerefer to grating of interest for range
Grating(s)Single, kinematic grating holder. Two grating turret optional.
SlitContinuously variable micrometer actuated width 0.01 to 1 mm. Settable height.
VacuumUltra High Vacuum 10E-9 torr standard
Focal Plane40 mm microchannel plate or 25 mm direct detection CCD


Performance with various diffraction gratings:

Grating Groove Density (g/mm)2105450290
Spectral Resolution (nm,FWHM)0.050.10.16
Focal Plane Width (mm)404040
Wavelength Range (nm)9.5 to 3210 to 11015.5 to 170
Energy Range (eV)130 to 39124 to 1180 to 7


上一篇:水位差仪:测量水位的高精度工具 下一篇:耐腐蚀水下电缆线,机器人柔性电缆线技术参数表现力
热线电话 在线询价
提示

仪表网采购电话