One meter Seya-Namioka Monochromator

One meter Seya-Namioka Monochromator

参考价: 面议

具体成交价以合同协议为准
2023-03-08 08:22:04
69
产品属性
关闭
北京埃光科技有限公司

北京埃光科技有限公司

免费会员
收藏

组合推荐相似产品

产品简介

Scanning and MCP or CCD SpectroscopyIncreased work spaceInternally welded Stainless steel constructionPrecision Micrometer Adjustable SlitsHigh resolution

详细介绍

商品描述

Scanning and MCP or CCD Spectroscopy

Increased work space

Internally welded Stainless steel construction

Precision Micrometer Adjustable Slits

High resolution



· Scanning and MCP or CCD Spectroscopy

· Increased work space

· Internally welded Stainless steel construction

· Precision Micrometer Adjustable Slits

· High resolution


The Model 231 1-meter focal length Seya-Namioka uses simple grating rotation to scan wavelengths. An angle of 70° 15' subtends the entrance and exit slit arms. The focal length provides excellent resolution and a lot of work space. Stainless steel construction makes this design suitable for use with high vacuum and contaminant free experimental chambers or microchannel plate intensified detectors. The McPherson Model 231 Seya-Namioka features fixed entrance and exit slit positions as well as fixed beam direction from the exit slit. When used as a scanning spectrometer these valuable characteristics make the Model 231 suitable for use in applications at synchrotron storage rings. The Model 231M4 (231 version or revision 4) was designed specifically synchrotron applications. It offers white light bypass optics as well as diverting optics when the use of the spectrometer is required. Multiple grating turrets are employed to further simplify instrument demands during experiments.


Model 231 PDF Data Sheet


Optical Design

McPherson Model 231 1-meter focal length f/23 Vacuum Monochromator

Angle between Slits

70° 15'

Focal Length

1-meter focal length Seya-Namioka design

f/no.

23

Wavelength Range

refer to grating of interest for range

Wavelength Accuracy

+/-0.1-nm (on counter, with 1200 G/mm grating)

Wavelength Reproducibility

+/- 0.005 nm (with 1200 G/mm grating)

Grating Size

30 x 50-mm ruled area

Focal Plane

Multiply dispersion by the width of your detector for range


Performance with various diffraction gratings:

Grating (G/mm) (others available)

2400

1800

1200

600

300

Wavelength Range from 30-nm to

150nm

200nm

300nm

600nm

1.2um

Resolution (nm) at 313.1-nm

0.013

0.018

0.025

0.05

0.10

Dispersion (nm/mm)

0.42

0.55

0.83

1.66

3.33

Blaze Wavelength: (nm)

Holo

Holo

Holo

Holo

Holo


80nm


150nm

150nm

700nm





160nm






276nm


* gratings work best from 2/3 to 3/2 the blaze wavelength