品牌
经销商厂商性质
上海所在地
Santec MPM-210H是一款的多通道,光器件IL、WDL及PDL测量的光功率计,包括高密度波分复用(DWDM),阵列波导光栅(AWG)及波长选择开关(WSS)等器件测试。通过搭配Santec TSL系列可调谐激光器,并带功率计输出,MPM-210H完成实时高精确IL测量。
参数 | 单位 | 规格书 | 备注 | |
模块型号 | – | Up to 5 | ||
外部接口 | 用于功率计 | – | GP-IB, Ethernet, RS-232C | |
用于系统 | – | USB | ||
触发输入 | – | TTL (3.3 V) | BNC | |
触发输出 | – | TTL (3.3 V) | BNC | |
功率监控 | V | 0 to 3 | BNC | |
供给电压 | V | AC 100 to 240, 50 / 60 Hz | ||
功耗 | VA | 50 | ||
环境温度 | deg | 10 to 40 | ||
环境湿度 | % | <> | non condensing | |
重 | kg | 6 | ||
体积 (W) x (D) x (H) | mm | 210 x 350 x 133 |
Parameter | Unit | MPM-211 | MPM-212 | MPM-215 *1 | Notes |
Sensor element | - | InGaAs | |||
Wavelength range | nm | 1250 to 1680 | |||
Specification wavelength range | nm | 1250 to 1630 | |||
Power dynamic range | dBm | - 80 to + 10 | - 70 to + 5 | ||
Dynamic range @Fixed gain (typ.) *2, 3 | dB | 45 | 70 | Logging mode | |
Number of gain range | - | 5 | 1 | ||
Maximum safe power | dBm | + 16 | |||
Total uncertainty | % | +/- 5 @- 60 to 9 dBm | +/- 5 @>-55 dBm | ||
Power resolution | dB | 0.001 | |||
Linearity *2, 4 | dB | +/- 0.03 @- 55 to 9 dBm | +/- 0.02 @>-40 dBm +/- 0.05 @>-50 dBm | | |
Polarization dependent responsivity (typ.) *2, 5 | dB | < 0.025 | @1525 to 1585 nm | ||
< 0.03 | @1270 to 1630 nm | ||||
Averaging time | sec | 10μ to 10 | |||
Data logging capability | – | 1,000,000 x 2 buffer per port | |||
Port number per module | – | 4 ports | 2 ports | 4 ports | |
Analog output *6 | None | With | None | ||
Connector type | – | FC |
参数 | 单位 | 新型号, MPM-213 | 备注 |
电流动态范围 | dBmA | - 70 to + 10 | 100 pA to 10 mA |
安全功率 | dBmA | 16 | |
总不确定度 | % | +/- 5 @- 45 to 9 dBm | 100 pA to 10 mA |
+/- 1 @>- 35 dBm (typ.) | Averaging time > 10 ms | ||
功率分辨率 | dB | 0.001 | |
线性度 | dB | +/- 0.03 | @- 45 to + 9 dBmA, Averaging time > 10 ms |
平均时间 | sec | 10μ to 10 | |
每个模块通道数 | – | 4 ports | |
数据记录能力 | – | 1,000,000 per port x2 | |
连接器类型 | – | BNC Connecter | |
反向偏置电压 | V | – | |
环境温度 | deg | 10 to 40 | |
环境湿度 | % | <> | 无冷凝 |
体积 (W) x (D) x (H) | mm | 30.3 x 183.5 x 114.8 | 不包括连接器 |
*1: This module is not compatible with other modules on a same Main Frame.
*2: Temperature: 23 +/- 5 ℃,
*3: Averaging time >50us, Within ±1 ºC after zeroing, At gain 1, 2 or 3 for MPM-211/212
*4: For MPM-211/212, Averaging time >100ms, Auto range mode, For MPM-215, Averaging time >30us
*5: SMF, Straight connector
*6: Output voltage 0 to 2 V, Measurement dynamic range > 30 dB (typ.)