PSP103型气相色谱仪
执行标准:GB/T 17623-1998 GB/T 7252-2001 DL/T 722-2000
PSP103型气相色谱仪是本公司研制的新一代气相色谱仪。仪器采用微机自动控制、全键盘操作、大屏幕液晶中文显示,具有控温精度高、性能稳定可靠、灵敏度高、重复性好等特点。
执行标准:GB/T 17623-1998《绝缘油中溶解气体组分含量的气相色谱测定法》
GB/T 7252-2001《变压器油中溶解气体分析和判断导则》
DL/T 722-2000《变压器油中溶解气体分析和判断导则》
仪器采用双柱并联分流系统。配有TCD检测器和双FID检测器以及甲烷转化器,能一次进样实现油中溶解气体九组分(H2、O2、N2、CH4、C2H2、C2H4、C2H6、CO、CO2)的全分析。可用于电力系统油气设备内部故障检测,氢冷发电机冷却介质分析、六氟化硫(SF6)杂质分析、锅炉烟气分析、天然气分析和环境监测分析等。既可作分析,又可作通用分析,因而成为石油、化工、矿山等系统作气体分析的效设备。
1、温度范围
柱室:室温加10℃~399℃
热导检测器:室温加10℃~300℃
氢焰检测器:室温加10℃~350℃
转化炉:室温加10℃~399℃
温度控制精度:±0.1℃
温度显示精度:0.1℃
2、TCD检测器
TCD灵敏度:≥5000mv.ml/mg(正十六烷/异辛烷)
噪声:≤0.02mV
漂移:≤0.05mV/30min
3、FID检测器
检测限:≤5×10-12g/s(正十六烷/异辛烷)
噪声:≤0.02mV
漂移:≤0.05mV/30min
4、定性重复性误差:≤1%
5、定量重复性误差:≤3%
6、低检测量
H2 :≤2ppm
CO、CO2:≤2ppm
烃类:≤0.1 ppm
The PSP103 Gas chromatograph
Implementation of the standards: GB / T 17623-1998 GB / T 7252-2001 DL / T 722-2000
PSP103 type gas chromatograph is the latest development of a new generation of gas chromatograph. The instrument uses a microcomputer automatic control, full keyboard, large screen LCD Chinese display with high precision temperature control, stable and reliable performance, high sensitivity, good repeatability.
Implementation of the standards: GB / T 17623-1998 "component content of dissolved gases in the insulating oil gas chromatographic method
GB / T 7252-2001 "transformer oil dissolved gas analysis and judgment guide
DL / T 722-2000 "transformer oil dissolved gas analysis and judgment Guidelines
The instrument uses a double column parallel shunt system. Equipped with a TCD detector and dual FID detector, as well as the conversion of methane, a single injection Dissolved Gas nine components (H2, O2, N2, CH4, C2H2, C2H4, C2H6, CO, CO2) of the full analysis. Power system can be used for oil and gas equipment internal fault detection, hydrogen-cooled generator cooling medium, and sulfur hexafluoride (SF6) impurity analysis, boiler flue gas analysis, gas analysis and environmental monitoring and analysis. Both exclusive analysis, but also for the general analysis, and thus become the most effective equipment for the petroleum, chemical, mining and other systems for gas analysis.
1, the temperature range
The column room: room temperature and 10 ℃ ~ 399 ℃
Thermal conductivity detector: room temperature and 10 ℃ ~ 300 ℃
Hydrogen flame detector: room temperature plus 10 ° C to 350 ° C
The conversion furnace: room temperature plus 10 ℃ ~ 399 ℃
Temperature control accuracy: ± 0.1 ℃
Temperature display accuracy: 0.1 ° C
2, TCD detector
TCD sensitivity: ≥ 5000mv.ml/mg (n-hexadecane / isooctane)
Noise: ≤ 0.02mV
Drift: ≤ 0.05mV/30min
3, FID detector
Detection limit: ≤ 5 × 10-12g / s (n-hexadecane / isooctane)
Noise: ≤ 0.02mV
Drift: ≤ 0.05mV/30min
Qualitative repeatability error: ≤ 1%
Quantitative repeatability error: ≤ 3%
6, the lowest detectable amount
H2: ≤ 2ppm
CO, CO2: ≤ 2ppm
Hydrocarbons: ≤ 0.1 ppm