Technical Data HLT 560, 230 V, 50 Hz
Ambient temperature 10-35
Backing pump Rotary vane pump 5 m3/h
Cold start to ready 3 min
Detectable gases 4He, 3He, H2
Flange (in) DN 25 ISO-KF
Inlet pressure max. 25 mbar
Input: Analog ActiveLine
vacuum
transmitter
Input: Digital Start/Stop, Vent,
Cero, Calibration,
Bypass detection
Interface RS-232-C, RS- 485
Leak rate display 10-12 - 1 mbar l/s
Output: Analog Leakage rate
signal 2 x 0-10 V,
lin/log
Output: Digital Ready to start,
Ready to
measure, Leak,
Error, Bypass
valve, Calibration
acknowledge
Power consumption 400 VA
Pumping speed for He 2.5 l/s
Pumping time to ready: 0.5 l test volume 1 s
Pumping time to ready: 100 l test volume 590 s
Pumping time to ready: 10 l test volume 45 s
Relay: Switching capacity 2 x 230 V AC
Relay: Switching current 3 A
Response time 0.5 s
Roughing pump no
Smallest detectable leak rate for 4He (according AVS 2.1) < 5?0-12 mbar l/s
Sniffer probe, detectable leak rate < 5?0-8 mbar l/s
Supply 230 V, 50 Hz
Test method Vacuum and
sniffing leak
detection
Weight 44 kg