Face Mask Bacterial Filtration Efficiency (BFE) Tester

AG-6956Face Mask Bacterial Filtration Efficiency (BFE) Tester

参考价: 面议

具体成交价以合同协议为准
2021-07-16 13:25:06
304
产品属性
关闭
深圳市安规检测设备有限公司

深圳市安规检测设备有限公司

免费会员
收藏

组合推荐相似产品

产品简介

A.Scope of Application:It is used for the performance test of the bacteria filtration efficiency of face masks by metrological verification departments, scientific research institutes, face mask ma......

详细介绍

A.     Scope of Application:

Itis used for the performance test of the bacteria filtration efficiency of facemasks by metrological verification departments, scientific research institutes,face mask manufacturers, etc.


B. Applicable Standards:

GB/T32610-2016: Technical specification of daily protective mask

YY0469-2011:Surgical mask


C. Equipment Features:

1.     Negativepressure experiment system to ensure the safety of operators.

2.     Thenegative pressure cabinet has a built-in peristaltic pump, and AB two-way SixStage Microbial Andersen Cascade Impactor.

3.     Theflow rate of the peristaltic pump can be set.

4.     Thespray flow of the bacterial liquid of the dedicated microbial aerosol generatorcan be set.

5.     Goodatomization effect.

6.     Embeddedhigh-speed industrial microcomputer control.

7.     10.4-inchindustrial-grade high-brightness color touch screen.

8.     USBinterface, support thumb drive data transfer.

9.     Built-inhigh-brightness lighting lamp.

10. AFront-opening glass door is convenient for experimenters to observe andoperate.

11. AC220/50HZ


D. Technical Parameters

Main Parameters

Range

Accuracy

Maximum Tolerable Deviation

Channel A Sample Flow

28.3L/min

0.1L/min

Better than ±2.5%

Channel B Sample Flow

28.3L/min

0.1L/min

Better than ±2.5%

Spray Flow

(810)L/min

0.1L/min

Better than ±5.0%

Peristaltic Pump Flow

(0.0063.0)mL/min

0.001ml/min

Better than ±2.5%

Aerosol Chamber Negative Pressure

(-90-120)Pa

0.1Pa

Better than ±2.0%

Temperature

(050)

Data Storage Capacity

>100000 sets

High-Efficiency Air Filter

Filtration efficiency ≥99.99% for particles over 0.3um

Aerosol Generator

Average particle diameter (3.0±0.3)μm, the geometric standard deviation ≤1.5

Particle Size Captured by two-way 6 Stage Microbial Andersen Cascade Impactor

Level I7μm, Level II (4.7~7)μm, Level III (3.3~4.7)μm, Level IV (2.1~3.3)μm, Level V (1.1~2.1)μm, Level VI (0.6~1.1)μm

Aerosol Chamber Dimensions

(Length 600 × diameter 85 × thickness 3) mm

Total particles before filtration

(2200±500)cfu

Cabinet Dimensions

(L 1200× W 700× H 500)mm

Overall Dimensions

(L 1600× W 600× H 1950+250)mm

Power Supply

AC220V±10%50Hz

Noise

<65dB(A)

Weight

About 150kg

Power Dissipation

<1500W