The thermal conductivity is a significant parameter for researching physical properties of a material. In aviation area, atomic energy area, and industry department of building material or non-metallic material, an actual measurement or a forecast is required for the related material. There are steady testing method and dynamic testing method. The equipment adopts modified guarded hot plate method which is in line with the national standard. The whole testing process is finished either automatic or manual which satisfied the requirement of a high precision test. Reference standard: GB/T 10294-2008.
Applications: The equipment applies to test the thermal conductivity of clintheriform homogeneous material under dry or different moisture conditions. Thermal conductivity range: 0.015-2 W/M*K
repeatability: ±2%
①Environment temperature: 10°~35℃
②Relative humidity: ≤80%RH
③Require to operate in the condition of a stable room temperature. The average daily temperature fluctuations ≤±1.5℃
Instrument consist of the built-in high precision regulated power supply, thermometer, hot plate heater, the main protection of hot plate, temperature measuring thermocouple, cold plate constant temperature water bath, computer testing system.
a. Specimen part: including constant temperature device, through room temperature to 80℃
b. Heating system: heater of guarded hot plate temperature and heater of measuring power.
c. Temperature measurement system: adopting the high precision digital display instrument to ensure the accuracy and stability, and realize zero internal compensation.
d. measuring power adopting constant heating and digital display, the accuracy is better than 1%.
e. computer test parts including touch screen, communications components, and software.
f. instrument panels and the various parts see figure 1:
(Figure 1)
1、Power switch 2、Work stop 3、Work start 4、Power light 5、Power regulation
6、Touch screen 7、Samples room 8、Cold plate lifting knob 9、Base seat
10、Rolling wheel