多气体培养箱

MG-70M-A多气体培养箱

参考价: 面议

具体成交价以合同协议为准
2021-09-25 10:50:56
279
产品属性
关闭
安捷来国际股份(香港)有限公司

安捷来国际股份(香港)有限公司

免费会员
收藏

组合推荐相似产品

产品简介

多气体培养箱:MG-70MIRCO2sensorandzirconiaO2sensorsrealizehighprecisedhypoxicculture

详细介绍

多气体培养箱:
MG-70M

IR CO2 sensor and zirconia O2 sensors realize high precised hypoxic culture.

MG-70M

Two gas cylinders connectable. Automatic cylinder changer are standardized in all models.

Each model can attach a shelf up to six.
(Images of 3 shelves attached)


Main Application


Mammalian cell culture
For research of iPS and cancer


Features

Possible highly precise hypoxic culture

This O2 sensor is not using conventional galvanic battery but zirconia battery which is stable for a long time and high sensitivity. It has a quick response and shortens the lead time of validation when installed and transition time to hypoxic condition.


●Quick transition to hypoxic condition

In MG-70M, the graph shows transition of DO inside PBS put on the same chamber until the DO inside the chamber goes up to set value.
It was verified that faster attainment than equivalent models of competitors(refer to values on each catalog) and its data agreed with theoretical values (O2 partial pressure between gas and liquid phase).


●Apoptosis data in hypoxic culture

PC12 cells to which apoptosis is induced by hypoxia condition. The hypoxic environment was confirmed by c*tion at 9 points in MG-70M.

Left panel: green arrowheads indicate apoptotic cells and white arrows indicate necrotic cells
Right panel: percentage of apoptotic and necrotic cells

Representative data were shown.


Dissolved oxygen (DO) database inside


It is installed of referencial data about DO concentration of several patterns inside the culturing containers when set at 37℃ .
By combination with non-invasive O2 sensor makes it possible to monitor DO, pH and CO2 in hypoxic culture.


Outstanding cost effectiveness and automatic cylinder changer

A c*tion performance is equivalent with other established companies with lower cost. For example, when two cylinders were connected and one side was consumed,it is automatically canged to new cylinder.
In 70M, CO2 and N2 (or O2) are connectable each two cylinders.


Specifications

Model MG-70M
Temperature Range Ambient+5ºC to +50ºC
Temp. Control Accuracy ±0.1 ºC
Heating System Heating method: airjacket
Humidification method: natural evaporation by trays
CO2 Sensor IR CO2 sensor
CO2 Range 0 to 20% (above atmosphere), Digital display, Increment: 0.1%
O2 Sensor Zirconia O2 sensor
O2 Range 1-25%(*1), Digital display
Dimensions 457×543×632Hmm
Inner Dimension Chamber:320×410×400Hmm
Capacity Capacity: 70L
Vessel availability: for example, 49 x φ35mm petri dish or 6 well plates or 14 x 25 cm2 TC per shelf.
Other Functions Hole for power supply cable etc. x1, Automatic change of cylinder, Date base of DO
Standard Accessory Shelf x 3

(*1)Depending on the conditions of CO2, it may not be stabilized in 18 - 22% of range. O2 gas cylinder is required for set above atmospheric concentration and N2 gas cylinder for set below atmosphere.

●Two levels stackable.
●MG-70M can attach a shelf up to six.

●MG-70M can be modified to cGMP standard.



上一篇:振荡培养箱异常噪音的排查与处理方法 下一篇:培养箱的类型与应用详解
热线电话 在线询价
提示

仪表网采购电话