Specifications |
Size Range: | 0.3 - 25.0μm |
Channel Sizes: | 0.3, 0.5, 1.0, 3.0, 5.0, 10.0μm 0.3, 0.5, 1.0, 5.0, 10.0, 25.0μm |
8-Channel: | 0.3, 0.5, 0.7, 1.0, 3.0, 5.0, 7.0, 10.0μm |
Flow Rate: | 1.0 CFM (28.3 LPM) |
Counting Efficiency: | 3100+, 50% @ 0.3µm; @ sizes > 0.45µm Per JIS B 9921, ISO 21501-4 |
Light Source: | Extreme Life Laser Diode Technology with >20 Year Mean Time To Failure (MTTF) |
Zero Count Level: | <1 count / 5 minutes (per JIS B 9921) |
Concentration Limits: | 500,000 particles / ft3 @ 5% coincidence loss |
Calibration: | NIST Traceable, per ISO 21501-4 |
Count Modes: | Manual, Automatic, Beep, Concentration and Real Time (with Let or LMS Express RT/RT+) |
Data Storage: | 3,000 Sample Records (Rotating Buffer) which includes particle and environmental data, locations and times |
Communication Modes: | RS-232 via RJ-45 to PC, RS-485/MODBUS |
Supporting Software: | LMS XChange, LMS Express, LMS Express RT/RT+, Let |
Environmental Sensors: | Optional (up to 4): Temperature/Relative Humidity, Air Velocity, Differential Pressure |
Touch Screen Display: | 5.7-inch (14.47 cm) color, 320x240 |
Printer: | Built-in Thermal Printer |
Reports: | ISO-14644-1, FS-209E, EU GMP Annex 1 |
Alarm: | Integrated, Adjustable Alarm and External Alarm Connection: Alarms on Counts and Low Battery |
Enclosure: | 304 Stainless Steel |
Sample Output: | Internally filtered to HEPA standards (>99.997% @ 0.3μm) |
Vacuum Source: | Internal flow-controlled pump |
Power: | 100-240 VAC, 50-60 Hz |
Battery: | Li-Ion, Removable & Rechargeable |
Dimensions: | 9.96” (l) x 8.0” (w) x 10.16” (h) [25.3 x 20.32 x 25.81 cm] |
Weight: | With Battery: 14.9 lbs (6.8 kg); Without Battery: 12.95 lbs (5.9 kg) |
Environmental Conditions |
Operating: | 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing |
Storage: | 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing |
Accessories |
Included: | Operating Manual on CD; Isokinetic Sample Probe with Tripod and Tubing; Purge Filter; Removable/Rechargeable Li-Ion Battery; Power Cable and Spare Fuse; LMS XChange Software; Printer; Printer Paper |
Optional: | Printed Operating Manual; Network Adapter; Wireless Network Adapter; Shipping Case with Wheels; Cleanroom Cart; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; Filter Scanning Probe; High Pressure Diffuser; Spare Removable/Rechargeable Li-Ion Battery; External Battery Charger; 32-port Universal Manifold System; LMS Express/RT/RT+ Software; Let Software; Validation Manual |