p-Dichlorobenzene | NO.127P |
C6H4Cl2 |
Calibration Range | 100-3000μg/m3 |
Measuring Range | 100-3000μg/m3 |
Sampling (flow) Rate | 100mL/min |
Sampling Volume | 3000mL |
Sampling Time | 30min |
Detecting Limit | 20μg/m3 (3000mL) |
Colour Change | yellow → Pale reddish purple |
Corrections for temperature/humidity | Necessary for 5 to 35 ºC Unnecessary for R.H.20 to 80% |
Shelf life | 2 years |
Substance(s) | Interference | Colour change by itself |
Ammonia | No | No | |
Hydrogen chloride, Chlorine | No | No | |
Nitrogen oxides | No | No | |
Vinyl chloride | + | Pale reddish purple | |
1,2-Dichloroethylene | + | Pale reddish purple | |
Trichloroethylene | + | Pale reddish purple | |
Tetrachloroethylene | + | Pale reddish purple | |
1,1,1-Trichloroethane | No | No | |
Aromatic hydrocarbons | No | No | |
Formaldehyde | No | No |
Calibration gas generation | Permeation tube method |
Electric sampling pump This detector tube is suited for measuring p-dichlorobenzene in indoor environments. The GSP-200 / GSP-300FT-2 series automatic gas samplers (or comparable sampling devices/pumps) should always be used for optimal performance. A deviation of +/-10% in the sampling flowrate (i.e. amount of air flowing through the tube at a given moment) or the sampling amount (total sampled amount of target substance) causes an average margin of error of approx. +/-10% (maximum +/-20%) in the indicated value. |
TLV-TWA | TLV-S | Explosive range |
10ppm | 〖〖 | 2.5 - 16% |