Toluene | NO.122P |
C6H5CH3 |
Calibration Range | 100-2500μg/m3 |
Measuring range | 100-7000μg/m3 |
Sampling (flow) Rate | 200mL/min |
Sampling Volume | 6000mL, 2000mL |
Sampling Time | 30min, 10min |
Detecting Limit | 50μg/m3 |
Colour Change | White → Pale brown |
Corrections for temperature/humidity | Unnecessary for 5 to 35 ºC Unnecessary for R.H.0 to 80% |
Shelf life | 2 years |
Substance(s) | Interference | Colour change by itself |
Ethanol | No | No | |
Xylene,Styrene | 1000μg/m3 | + | Pale brown |
p-Dichlorobenzene | No | No | |
α-Pinene | No | No | |
n-Hexane | No | No | |
Aromatic hydrocarbons | + | Pale brown | |
Formaldehyde | 1/3 | No | No |
Substance(s) | Correction Factor | Pump Strokes [n] | Measuring Range |
Ethyl benzene | coefficient: 1.1 | 200mL/min x 30min | 110-2750μg/m3 |
Xylene | coefficient: 5.4 | 200mL/min x 30min | 540-13500μg/m3 |
Calibration gas generation | Permeation tube method |
Electric sampling pump This detector tube is suited for measuring toluene in indoor environments. The GSP-200 / GSP-300FT-2 series automatic gas samplers (or comparable sampling devices/pumps) should always be used for optimal performance. A deviation of +/-10% in the sampling flowrate (i.e. amount of air flowing through the tube at a given moment) or the sampling amount (total sampled amount of target substance) causes an average margin of error of approx. +/-10% (maximum +/-20%) in the indicated value. |
TLV-TWA | TLV-S | Explosive range |
50ppm | -- | 1.4×6.7% |