GF+SIGNET 2720系列前置放大器
+GF+SIGNET 2720系列前置放大器 (已停产,替换型号为2760系列)
3-2720 前置放大器
Signet2720前置放大器是应用中所必须的,它可以防止公共工业环境中的各种电磁噪音,对PH/ORP电极有较微弱的输出信号的干扰。Signet2720旋转锁前置放大器,在传感器测量点的安装的。它就*大限度地保护长度可达120m的加长电缆。的旋转锁设计简化了安装和维护,它能使电极能快速连接和拆卸。放大器的CPVC本体,在浸入式和在线应用中都不易破损,延长了工作寿命。如果需要远程前置放大器(如:罐温>80℃)或接其他生产商的电极,可用Signet2721远程前置放大器。
2720 PH/ORP前置放大器的特点与性能:
·兼容的Signet pH/ORP探头:3-2714、3-2715、3-2716、3-2717
·信号转换:mV信号1:1增强
·前置放大器输出信号:mV信号
·安装方式:在线式 或 沉入式
·壳体材料:CPVC
·*大工作温度:85℃
·能够兼容的Signet变送器:3-8750
技术数据
输入范围:±2500mV
温度:0℃——80
℃
*大压力/温度等级:7bar max @ =<65
℃
4bar max @ =<80℃
需要的电源:±4.5-±8VPC双路供电(由Signet PH/ORP仪表提供)
*大电流:〈1mA,双路供电
温度接触电阻:〈0.1Ω
终端连接器:旋转锁接Signet271X电极
输入阻抗:〉1011Ω 材料:CPVC
电缆:6芯金属屏蔽24AWG
电缆长度:4.5m 标准(*大120m)
标准论证:
CE UL 制造符合ISO9001:2000质量 ISO14001:2004环境管理标准
+GF+SIGNET 2720系列前置放大器详细英文介绍
| 2720/2721 Preamplifier | | Pre-amplification is required to protect the output signal of pH & ORP electrodes from a wide variety of electromagnetic interference common in industrial environments. The Signet 2720 Preamplifier is installed locally, at the sensing point, to maximize this protection while allowing cable extensions up to 120 m (400 ft). Signet's unique twist-lock design simplifies installation and maintenance by providing for the quick connection and disconnection of electrodes. And the CPVC body is virtually unbreakable for extended service life in both submersible and in-line applications. If remote pre-amplification is necessary (e.g., tank temperature > 80° C), or if connecting to another manufacturer's electrode, use the Signet 2721 Remote Preamplifier. Datasheet | | | English | German | French | | 2720 Operation Manuals | English | Danish | French | German | Portuguese | Spanish | Italian | | 2721 Operation Manuals | English | | | | | |
| Features | | - Twist-lock design allows quick connect/disconnect of sensors
- Local pre-amplification for submersible and in-line installations
- Unbreakable CPVC body
- Designed for use with Signet pH & ORP Twist-lock Electrodes to assure signal integrity up to 120 m. (400 ft.)
| | | | | | Applications | | - Water & Wastewater Treatment
- Neutralization Systems
- Scrubber Control
- Effluent Monitoring
- Surface Finishing
- Flocculent Coagulation
- Heavy Metal Removal and Recovery
- Toxics Destruction
- Sanitization Systems
- Commercial Pools & Spas
- Aquatic Animal Life Support Systems
- Process Control
| | | | | | Compatible Products | | - 2714-2717 pH & ORP Electrodes
- 5700 ProPoint pH/ORP Monitor (2760 only)
- 8750 ProcessPro pH/ORP Transmitter (2760 only)
| | |