ThermoTrans 210/211ThermoTrans 210/211 德国(KNICK)科尼克 温度变送器
品牌
经销商厂商性质
所在地
WP101温度变送器(上润仪表)
面议WP201/202系列隔离转换模块(香港上润)
面议智能数学运算器(上润仪表)
面议WP30电压/电流/配电隔离转换模块(上润仪表)
面议虹润HR-WP-N20热电偶/热电阻温度变送模块(虹润仪表) 隔离器
面议虹润HR-WP-N20智能配电模块(虹润仪表) 隔离器
面议虹润HR-WP-N20智能电压/电流变送器(虹润仪表) 隔离器
面议虹润HR-WP-N20智能数学运算模块(虹润仪表) 隔离器
面议虹润HR-WP-N20智能频率转换模块(虹润仪表) 隔离器
面议虹润HR-WP-N20温湿度隔离传感器/变送器(虹润仪表)
面议虹润HR-WP-N40热电偶、热电阻温度变送模块(虹润仪表) 隔离器
面议虹润HR-WP-N40频率变送模块(虹润仪表) 隔离器
面议ThermoTrans 210/211 德国(KNICK)科尼克 温度变送器
隔离变送热电隅信号、输出0...20mA,4...20mA,0...10V,工作电压1000V,测试电压4kV。
Input data | Sensor type | Range | |||||||||||
Type B | DIN / IEC 584-1 | 0 … | +1820 °C | ||||||||||
Type E | DIN / IEC 584-1 | –270 … | +1000 °C | ||||||||||
Type J | DIN / IEC 584-1 | –210 … | +1200 °C | ||||||||||
Type K | DIN / IEC 584-1 | –270 … | +1372 °C | ||||||||||
Type L | DIN 43710 | –200 … | +900 °C | ||||||||||
Type N | ASTM E 230-87 | –270 … | +1300 °C | ||||||||||
Type R | DIN / IEC 584-1 | –50 … | +1767 °C | ||||||||||
Type S | DIN / IEC 584-1 | –50 … | +1767 °C | ||||||||||
Type T | DIN / IEC 584-1 | –270 … | +400 °C | ||||||||||
Type U | DIN 43710 | –200 … | +600 °C | ||||||||||
Voltage input | –20 … +100 mV | ||||||||||||
Input resistance | >10 Mohms | ||||||||||||
Span (configurable) | Min. ≥2 mV, max. measuring end – measuring start | ||||||||||||
Sensor failure monitoring | All inputs for open circuit (not with voltage measurement) | ||||||||||||
Input error limits | ±10 µV +0.05 % of meas. val. | ||||||||||||
Temperature coefficient | 0.01 %/K full scale (average TC in permitted operating temperature range, | ||||||||||||
at input | reference temperature 23 ºC) | ||||||||||||
Reference junction input | Internal Pt 100 | <±1.0 K | |||||||||||
(configurable) | External Pt 100 | <±0.3 K + error of Pt 100 used |
Output data | ||||||||||
Output signal | Model 210: 0 / 4 ... 20 mA, impressed current, load voltage ≤10 V | |||||||||
(0 … 100 %) | Model 211: 0 ... 10 V, impressed voltage, load current ≤10 mA | |||||||||
Resolution | Approx. 8000 steps (for 0 ... 100 %) | |||||||||
Control range | –2.5 % ... 102.5 % of span | |||||||||
Overload range | Model 210: –1.0 mA or 22 mA | |||||||||
with error message | Model 211: –0.5 V or 11 V | |||||||||
Output error limits | 0.1 % full scale | |||||||||
Temperature coefficient | 0.01 %/K full scale (average TC in permitted operating temperature range, | |||||||||
at output | reference temperature 23 ºC) | |||||||||
Residual ripple at output | <10 mVpp + digitalization error of input |
Transmission behavior | ||||||||||
Characteristic | Temperature-linear, voltage-linear or customer-specific rising or falling | |||||||||
Measuring rate | Approx. 1 / s | |||||||||
Response time T99 | ≤900 ms | |||||||||
Digital output filter | T99 = 0 ... 100 s (1st order filter) |
Power supply | ||||||||||
Power supply | 230 V AC –15 % +10 %, 48 … 62 Hz, approx. 2 VA | |||||||||
Option 336: | 24 V AC / DC | AC: –15 % +10 %, 48 … 500 Hz, approx. 1.5 VA | ||||||||
DC: –15 % +20 %, approx. 1.2 W | ||||||||||
Option 363: | 115 V AC –15 % +10 %, 48 … 62 Hz, approx. 2 VA |
Isolation | ||||||||||
Galvanic isolation | 3-port isolation between input, output and power supply | |||||||||
Test voltage | 4 kV AC (input against output and power supply) | |||||||||
3 kV AC (output against power supply) | ||||||||||
Working voltage | 1000 V AC/DC input against output and power supply with overvoltage category II | |||||||||
(basic isolation) | and pollution degree 2, | |||||||||
330 V AC/DC output against power supply with overvoltage category II and | ||||||||||
pollution degree 1 according to EN 61010-1 | ||||||||||
For applications with high working voltages, you should ensure there is sufficient | ||||||||||
spacing or isolation from neighboring devices and protection against electric shocks. | ||||||||||
Protection against electri | Safe isolation according to EN 61140 by reinforced insulation in accordance with | |||||||||
shock | EN 61010-1. | |||||||||
Working voltages with overvoltage category II and pollution degree 2: | ||||||||||
600 V AC / DC for input against output and power supply | ||||||||||
with overvoltage category II and pollution degree 2: | ||||||||||
300 V AC / DC for output against power supply. | ||||||||||
For applications with high working voltages, you should ensure there is sufficient | ||||||||||
spacing or isolation from neighboring devices and protection against electric shocks. |
Standards and approvals | ||||||||||
Surge withstand | 5 kV 1.2 / 50 µs according to IEC 255-4 | |||||||||
EMC 1) | EMC directive 89/336/EEC, EN 61326, NAMUR recommendation NE 21 |
Other data | ||||||||||
Interface | Optical, interface adapter on RS 232 interface (PC) is included in the | |||||||||
(only user-definable models) | ZU 0254 communications kit | |||||||||
Ambient temperature | Operation: | –10 ... +60 ºC | ||||||||
Transport and storage: | –30 ... +80 ºC | |||||||||
Housing | Modular housing A7, width 22.5 mm, screw terminals | |||||||||
See dimension drawings for further dimensions | ||||||||||
Protection class | Housing IP 40, terminals IP 20 | |||||||||
Mounting | With snap-on mounting for 35 mm top hat rail according to EN 50022-35, | |||||||||
width 22.5 mm, See dimension drawings for conductor cross section | ||||||||||
Weight | Approx. 300 g | |||||||||
1) Slight deviations are possible while there is interference from RF radiation |