品牌
其他厂商性质
所在地
全自动运动粘度仪/Automatic kinematic viscometer
面议赛波特粘度计Saybolt viscometer
面议赛波特粘度计Saybolt Viscosity Bath
面议气泡粘度计 Bubble Viscometers
面议英国 Hydramotion海默绅 VL7-100B-d21便携式粘度计
面议BU粘度单位 淀粉粘度计
面议博勒飞Brookfield DV-1 粘度计
面议BYK Gardner 气泡粘度计空玻管1
面议TQC Rotothinner旋转粘度计
面议荷兰TQC Krebs克雷布斯粘度计
面议马康 malcom PCU-02V 微量螺旋式粘度计
面议malcom 马康 PC-10 粘度计
面议美国加德纳 粘度杯种类和量程:
1)EZ系列蔡恩杯
Cup Number | Seconds Range | Centistoke Range | Midrange Sensitivity¹ | Calibration Oil Number² |
1 | 40 to 60 | 10 to 36 | 1.3 | G-10/19 |
2 | 20 to 60 | 19 to 156 | 3.3 | G-60/117 |
3 | 12 to 60 | 64 to 596 | 10.5 | G-200/458 |
4 | 10 to 60 | 79 to 784 | 13.9 | G-200/458 |
5 | 10 to 60 | 161 to 1401 | 24.2 | G-350/878 |
2)S90系列蔡恩杯
Range Specification | |||
Item Number | Cup No. | Seconds Range | Centistoke Range |
VI-2101 | 1 | 31 to 60 | 15 to 78 |
VI-2102 | 2 | 19 to 60 | 39 to 238 |
VI-2103 | 3 | 11 to 60 | 63 to 604 |
VI-2104 | 4 | 10 to 60 | 97 to 899 |
VI-2105 | 5 | 10 to 60 | 219 to 1627 |
3)福特杯
Technical Data | |||
Ford Standard Cup Number | Conversion Chart Seconds Range | Conversion Chart Centistoke Viscosity Range | Calibration Liquid |
0 | 57 to 120 | 2 to 14 | G-6 |
1 | 60 to 120 | 12 to 40 | G-10 |
2 | 37 to 120 | 25 to 120 | G-35 |
3 | 23 to 120 | 29 to 273 | G-60 |
4 | 16 to 120 | 34 to 441 | G-60 |
5 | 10 to 120 | 77 to 1413 | G-200 |
4)ISO杯
Technical Data | ||||
Orifice | Seconds Range | Centistoke Range | Midrange Sensitivity ¹ | Calibration Oil No./ Cenistokes ² |
3mm | 27 → 120 | 4.6 → 52 | .05 | #G 10/19Cs |
4mm | 21 → 120 | 19 → 163 | 1.4 | #G 35/68Cs |
6mm | 17 → 120 | 84 → 823 | 7.1 | #G 100/233Cs |
8mm | 10 → 120 | 187 → 2611 | 22 | #G 350/878Cs |
¹ Stated as centistokes per second of efflux time. ² Centistoke values are nominal - actual values printed on labels |
5)DIN杯
Technical Data Cup-Orifice | |
Orifice Diameter (Nominal): | 4mm |
Efflux Time Seconds Range: | 15 to 20 |
Centistoke Range: | 38 to 545 |
Midrange Sensitivity: (Centistokes per Second) | 4.7 |
Standard Calibrating Oil: | G100 |
Viscosity of Standard (Nominal Cs. - Use label value: | 233 |
6)费希尔杯(Fisher杯)
Technical Data | ||||
Orifice No. | Seconds Range | Centistoke Range | Midrange Sensitivity¹ | Calibration Oil No./Centistokes² |
1 | 22 - 60 | 11 - 48 | 0.9 | G 20 / 35 |
2 | 14 - 60 | 19 - 136 | 2.5 | G 35 / 68 |
3 | 10 - 60 | 35 - 320 | 10.3 | G 60 / 117 |
4 | 10 - 60 | 167 - 1125 | 19.0 | G 200 / 468 |
¹ Centistoke Values are Nominal - Actual Values Printed on Bottle Labels. ² Stated as Centistokes per Second of Efflux Time |
7)壳牌杯(停产中)
可订其他品牌
8)帕林杯(Parlin Cups)
echnical Data | ||||
Cup Number | Seconds Range | Centistoke Range | Midrange Sensitivity¹ | Calibration Oil Number/ Centistokes² |
1 | 25 to 110 | 7 to 165 | 1.6 | G-35-P/64 |
2 | 25 to 120 | 117 to 582 | 4.8 | G-100-P/270 |
3 | 25 to 110 | 498 to 2297 | 20.6 | G-350-P/1020 |
4 | 25 to 110 | 3445 to 15408 | 138.9 | G-J3000-P/10800 |
¹ Stated as Centistokes per second of efflux time ² Centistoke Values are nominal - actual values printed on labels |
9)斯坦霍尔粘度杯(Stein Hall Viscosity Cup)
The Stein Hall Viscosity Cup is used in viscosity testing where the number of seconds required for a set volume of adhesive to flow through the viscometer’s orifice. Viscosity is an important property of starch based adhesives and strongly affects bonding at the corrugator. Some pastes are too thick for certain paper and some are too thin, this is where viscosity testing is important. The apparent viscosity of adhesives used in the manufacture of corrugated containerboard is measured by the number of seconds required for 100 cc of adhesive to pass through an orifice in a disc. The instrument used throughout the industry since 1934 is called a Viscosity Cup or Stein-Hall Cup. The readings obtained do not directly correspond with any other viscosity scale.
The cup must be calibrated before performing the test. Check the orifice to make sure it is not plugged, corroded or worn. Manufacturing variations in orifice size and tube diameter make it necessary to calibrate each cup to a known standard. In addition, calibration may be required due to a worn disc or when replacing the disc.