The Nova 314 Portable Oxygen Analyzer is designed to measure the oxygen concentration of a clean, dry sample gas. In operation, a built-in sample pump draws the sample gas into the analyzer and through the sample conditioning filters, flow meter and O2 sensor. An external gas scrubber is available to remove any corrosive gases prior to entering the analyzer, if required.
The zirconia sensor is mounted in a flow-through block through which the sample gas flows. The sensor produces a non-linear µA output proportional to the oxygen concentration. This non linear signal is then linearized and amplified by an O2 processor board. The output is then directed to a main amplifier board which then connects to the digital readout meter and optional recorder output terminals.
Specifications
Method of Detection
Temperature controlled zirconium oxide sensor in amperometric mode
For continuous high accuracy analysis from PPM to 96% Oxygen (O2) in process gas streams. Other applications include combustion air O2 enrichment, flue gas analysis, O2 deficiency analysis, and many more. Note: Sample gas must not contain combustible gases or vapors.
Features
High accuracy, long-life zirconia O2 sensor
Rugged design that is easy to operate
Fast warm-up and response
Digital readout meter with backlight
Modular layout that is easy to maintain
Rechargeable battery operation
Built-in sample pump, filter, and flow meter
Bench Top cabinet with front-mounted controls and displays
Options
Recorder output of 0-1 V or 4-20 mA
Sample pre-cooler for hot samples
Condensate removal for wet sample gas
Suitcase (K) style cabinet available
AC power only operation
Low flow and O2 alarms
Detachable/portable datalogger
Calibration
Analyzed calibration gas of O2-free nitrogen (N2) for zero
Air @ 20.9% O2 or analyzed calibration gas of O2 in N2 for span
Unique Applications
The 314 series analyzers should not be used on sample gases containing flammable gases. These gases will be burned by the zirconia sensor along with any available O2 and cause the analyzer to display readings that are lower than the actual values.
All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.