Series 3000 Trace Oxygen (O2) Analyzer / Sensor

Series 3000 Trace Oxygen (O2) Analyzer / SensorSeries 3000 Trace Oxygen (O2) Analyzer / Sensor

参考价: 面议

具体成交价以合同协议为准
2023-03-07 10:33:11
107
产品属性
关闭
上海美盛自动化设备有限公司

上海美盛自动化设备有限公司

免费会员
收藏

组合推荐相似产品

产品简介

Series 3000 Trace Oxygen (O2) Analyzer / Sensor3000系列微量氧分析仪是坚固耐用,微处理器控制的设计用于连续或间歇性的微量氧气浓度的测量,3000系列微量氧分析仪是在单个或三个范围可从0-1到0-10,000每百万(百万分之一)PPM的测量分析

详细介绍

Series 3000 Trace Oxygen (O2) Analyzer / Sensor
3000系列微量氧分析仪是坚固耐用,微处理器控制的设计用于连续或间歇性的微量氧气浓度的测量,3000系列微量氧分析仪是在单个或三个范围可从0-1到0-10,000每百万(百万分之一)PPM的测量分析
  • Features and Benefits
  • System Description
  • Specifications
  • Explosion Proof Version
  • Applications
  • Dimensions
  • Download PDF Data Sheet
    (Requires Adobe Acrobat Reader)
  • Contact Us
Features Benefits

Safe to use

Some oxygen sensors must be initially filled with potassium hydroxide (caustic), a hazardous material. The Series 3000 Trace Oxygen Sensor is sealed; there is no such safety hazard.

Economically priced

Affords savings up to 50% and more over other electro- chemical oxygen analyzers.

Minimum maintenance

The Series 3000 Sensor never requires routine addition of water or potassium hydroxide. Operating costs are reduced.

Full sensor warranty Ensures cost-effective operation. Beware of "limited" or "pro-rated" sensor warranties. Warranty replacement of one of our competitor's sensor can exceed $3,000.

Visual and audible alarms

Ensures the highest level of protection.

Optional RS-232C or RS-485 communications Control several monitors on one communications channel along with security code protection.

System Description
Featured with the instrument is an advanced electrochemical oxygen sensor that provides exceptional performance, accuracy, and stability. Its enhanced mechanical design ensures longer sensor life, and virtually eliminates leakage of electrolyte, a nagging problem with wet cells.

Learn about about sample system handling for oxygen and carbon dioxide analyzers

The eloquence of the Series 3000 Trace Oxygen Analyzers are their ease of use. The Series 3000 Trace Oxygen Analyzers are equipped with three oxygen alarms and one status alarm relay, all of which can be configured for fail-safe operation. In addition to the alarm relays, each instrument is equipped with audible and visual alarm indicators. Available in a wide number of mechanical configurations from explosion proof (NEMA 7) to watertight (NEMA 4), the Series 3000 Trace Oxygen Analyzer affords maximum flexibility to meet the demands of the process control market. The Series 3000 Trace Oxygen Analyzer is priced thousands of dollars less than some other manufacturers. So before making that purchase decision, we would be delighted to submit a quotation for your consideration.
Back To Top

Specifications
PERFORMANCE
Measurement Ranges (parts per million)

Single-range
0-10,000
0- 5,000
0-1,000
0-100
0- 50
0-10

Three-range
0-100/1,000/10,000
0- 50/500/5,000
0-10/100/1,000
0- 5/50/500
0-1/10/100

Accuracy¹: ±1% of full scale (±5% fs 0-1ppm range)
Linearity: ±1%
Response Time: 90% of full scale response in less than 10 seconds (typical). Recovery time for ranges <10 ppm will depend on the prevailing conditions.
Sensor Type: Patented Ambient Temperature Electrochemical Sensor
Temperature Compensation: Standard
Operating Temperature:

  • 40° to 100° F (5° to 38° C)
  • <40° F use heated sensor enclosure
  • >100° F cooling of sensor required

Warranty: 3 years electronics,1 year sensor

ELECTRICAL
Display: 0.4" (10.2 mm) high, 4-1/2 digit LCD
Resolution:

  • 0.1 ppm for ranges <100 ppm
  • 1.0 ppm for ranges >100 ppm

Input Power: 115/230 VAC, 50-60Hz or 18-32 VDC. NICAD battery backup available as an option for some models.
Outputs: 4-20 mADC and 0-2 VDC standard. RS-232C or RS-485 optional. TTL output with 3 range instruments.
Oxygen Alarm Relays: Three (3) SPDT Form C contacts rated 10 A @ 30 VDC/115/230 VAC. User selectable for fail-safe or non fail-safe.
Instrument Status Alarm: Rated identically to the O2 relays
Audible Alarm: Internal buzzer with audible cancel

SAMPLE GAS CHARACTERISTICS
Flow Rate:
1.0 to 2.0 SCFH (0.5 to 1.0 liter/min)
Sample Gas Temperature: 32° to 122° F (0° to 50° C)
Sample Gas Pressure: 0.1 to 1.0 psig (1.04 kg/cm² to 1.10 kg/cm²)
Entrained Solids:

  • <3 mg/ft³: no in-line filter required
  • >3 mg/ft³: in-line filter is required

Hydrocarbon Mist:

  • <0.7 mg/ft³: no in-line filter required
  • >0.7 mg/ft³: in-line filter is required

CONSTRUCTION²
Electronics Control Unit (Benchtop):
Polycarbonate with a hinged clear front cover, rated NEMA 4X (IP 66)
Control Unit Dimensions:

  • 9.45 in. (240.5 mm) length
  • 6.50 in. (165.1 mm) width
  • 6.20 in. (157.5 mm) depth

Gas Connections: 1/4" compression fittings
Sensor Mounting: Local or optional remote mounting
Weight: 11lbs (4.98 kg) (Benchtop Config.)

¹ Stated at constant temperature and constant pressure.
² Other mechanical configurations available-consult factory.
³ Alpha Omega Instruments reserves the right to change or modify its product specifications without notice.
Back To Top

Learn more about oxygen sensor types

Explosion Proof Version

Series 3000 Trace Oxygen Analyzer with Electronics and Sensor in an Explosion Proof (NEMA 7) Enclosure

Series 3000 Trace Oxygen Analyzer with Electronics in a Panel Mounted Enclosure and Sensor in a Remote Explosion Proof (NEMA 7) Enclosure
Learn more about oxygen analyzer enclosure types

Back To Top

Applications

  • Ammonia Production
  • Battery Manufacturing
  • Bio-Research
  • Blanketing of Semiconductor Wafers
  • Bright Annealing
  • Ceramic Component Mfg.
  • Chemical Manufacturing
  • Contact Lens Manufacturing
  • Copper Thick Film Furnaces
  • Electron Beam Curing
  • Glass Manufacturing
  • Glove Boxes
  • Graphite Production
  • High Purity Gas Manufacturing
  • Inert Gas Generators
  • Light Bulb Manufacturing
  • Metal Processing
  • Petrochemical Manufacturing
  • Pipeline Leak Detection
  • Protection of Reactor Catalysts
  • Rubber Polymerization
  • Semiconductor Manufacturing
  • Sintering of Powdered Metals
  • Steel Production
  • Synthetics Manufacturing
  • Tritium Recovery
  • Vessel Blanketing
  • UV Curing
  • Welding of Critical Components

Back To Top


Dimensions

上一篇:上海理涛土工布动态穿孔测试仪:精准检测,助力工程品质提升 下一篇:什么叫移液器,移液器究竟是如何使用的呢?
热线电话 在线询价
提示

仪表网采购电话