MAIN UNIT |
Measured and Calculated Values | Dew point °C & °F |
Conversions | Automatic Pressure Compensated Conversions ppmV for Natural Gas and Ideal Gas Lb/mmscf and mg/Nm-³ for Natural Gas Dew point at a pressure input by the User for Natural Gas and Ideal Gas Calculations for natural gas moisture content based on either ISO 18453 or IGT#8 to customer order preference |
Keypad/Display | Capacitive touch-screen through glass Display: Vacuum Flourescent |
Data Logging | A rolling maximum of 150 data logs is available. Each log records time, date, moisture and pressure values for each channel. 5 minutes minimum and 60 minutes maximum logging intervals can be set by the user |
Communication & Outputs | Four non-isolated 4-20 mA channels; 500 Ω maximum load. Range and value settable by the user Modbus RTU @ 9600 baud-rate. |
Alarms | Two volt free contacts per channel; one process value and one instrument status. Also available via Modbus communications |
Main Unit Enclosure | Internal temperature control for condensation protection and stable analysis conditions |
Type | Flameproof EExd |
Construction | Body and lid cast Copper free Aluminum LM25(EN AC-42000). 15mm thick heat resistant glass window. Cable gland entries ½" NPT (female) |
Finish | Chromate primer, Polyester P9010 powder coated (Black). Meets BS3900 |
Electrical Connections | 2.5mm² two part terminals |
Hazardous Area Certification | ATEX Certification Code: Group II, Category 2 G |
Code | EEx d IIB + H2 T4 Tamb -40 +60°C (-40 +140°F) |
Ingress Protection | IP66 (NEMA 4) |
Measurement Range | -120°C dp to +30°C dp (-184 / +86°F); 1 ppbV to 30,000 ppmV |
Calibration Range | -100°C dp to + 20°C dp (-148 / + 68°F); 10 ppbV to 23,000 ppmV |
Accuracy | Dew point: ±1°C between -59.9 & + 20°C dp (±1.8°F: -75.9 to +68°F dp) Moisture content: ±10% of reading Dew point: ±2°C between -60 & -100°C dp (±3.6°F: -76 to -148°F dp) Moisture content: ±20% of reading Analysis Pressure: ±0.25% FS |
Measured Resolution | 0.1°C: -80 to + 20°C dp (0.2°F: -112 to + 68°F dp) 1°C: -100 to -80°C dp (2°F: -148 to -112°F dp) |
Displayed Resolution | Dew point: 0.1°C (0.2°F) Lb/mmscf: >999.9 = 0.1, <999.9 = 0.02 ppmV & mg/Nm-³: <100 = 0.001; >100 <1000 = 0.01; >1000 = 0.1 MPa & Barg: 0.1 p.s.i.g.: 1.0 |
Temperature Coefficient | Algorithm compensation |
Operating Pressure | Max. 206 Barg , (2987 p.s.i.g.) |
Sample Gas Connections | ¼" NPT (female) |
Sample Flow Rate | 0.0-0.4 m³/h (0-6.66 l/min) |
Power Supply | 90-264 VAC, 47 to 63 Hz, 110 to 300 VDC, 180 W |
Operating Environment | Indoor/Outdoor: -20°C to +60°C (-4 + 140°F) Shaded location: Max 95% RH |
In/Out Port Connections | ¼" NPT (female) |
Remote Interface Optional, for indoor safe area installation |
Host Capacity | Up to 31 Michell EExd Process Instrumentation Range analyzers in any combination of models via Modbus |
Display | Backlit color 15cm/6" LCD |
User Interface | Integral mouse, on-screen keyboard |
Functions | View and modify analyzer operating settings Record parameters to file and view virtual chart recorder Analyzer diagnostics |
Ports | LAN, USB2 |
Data Download | CSV (comma separated variable) format for easy import to Excel™ |
Remote Access | OPC (OLE for Process Control) over a LAN Imbedded web page/unique IP address for PC access |
Power Supply | 100-280 V, 50/60 Hz |