DSCUSB Advanced User Manual 一、Commands in Access OrderASCIInamedescriptiondatatypeaccessCMVVTempCompensatedmV VfloatROSTATStatus(Seetablelater)intROMVVFi

详细介绍

一、Commands in Access Order


ASCII name

description

datatype

access

CMVV

Temp Compensated mV/V

float

RO

STAT

Status (See table later)

int

RO

MVV

Filtered & factory calibrated mV/V

float

RO

SOUT

selected output

float

RO

SYS

main output

float

RO

TEMP

temperature

float

RO

SRAW

raw system output

float

RO

CELL

cell output

float

RO

FLAG

error flags (See table later)

int

RW

CRAW

raw cell output

float

RO

ELEC

electrical output

float

RO

SZ

system zero

float

RW

SYSN

snapshot result

float

RO

PEAK

Peak value

Float

RO

TROF

Trough value

Float

RO

CFCT

Communications failure count

Float

RW

VER

software version

byte

RO

SERL

serial number low

int

RO

SERH

serial number high

int

RO

STN

station number

int

RW

BAUD

baud rate select

byte

RW

OPCL

Output Control (Value select)

byte

RW

RATE

reading rate select

byte

RW

DP

digits after point

byte

RW

DPB

digits before point

byte

RW

NMVV

Nominal mV/V for scaling ELEC

float

RW

CGAI

cell gain

float

RW

COFS

cell offset

float

RW

CMIN

cell range min

float

RW

CMAX

cell range max

float

RW

CLN

lin n-points

byte

RW

CLX1..7

lin raw-values

float

RW

CLK1..7

lin corrections

float

RW

SGAI

system gain

float

RW

SOFS

system offset

float

RW

SMIN

system range min

float

RW

SMAX

system range max

float

RW

FFLV

Dynamic Filter Level

Float

RW

FFST

Dynamic filter steps

Float

RW

RST

reboot

X

SNAP

take snapshot

X

RSPT

Reset peak & trough

X

SCON

Shunt cal ON

X

SCOF

Shunt cal OFF

X

OPON

Digital Output on

-

X

OPOF

Digital output off

-

X

CTN

tempco n-points

byte

RW

CT1..5

tempco Temp points

float

RW

CTG1..5

tempco gain-adjust

float

RW

CTO1..5

tempco offset-adjust

float

RW


二、The various bits in the STAT value are as follows


Bit

Value

Description

Name

0

1

Setpoint output status

SPSTAT

1

2

Digital Input status (DSC ONLY)

IPSTAT

2

4

Temperature under range (TEMP)

TEMPUR

3

8

Temperature over-range (TEMP

TEMPOR

4

16

Strain gauge input under-range

ECOMUR

5

32

Strain gauge input over-range

ECOMOR

6

64

Cell under-range (CRAW)

CRAWUR

7

128

Cell over-range (CRAW)

CRAWOR

8

256

System under-range (SRAW)

SYSUR

9

512

System over-range (SRAW)

SYSOR

10

1024

(unused – reserved)

Unused

11

2048

Load Cell Integrity Error (LCINTEG)

LCINTEG

12

4096

Shunt Calibration Resistor ON

SCALON

13

8192

Stale output value (previously read)

OLDVAL

14

16384

(unused – reserved)

Unused

15

32768

(unused – reserved)

Unused


三、he various bits in the FLAG value are as follows

Bit

Value

Description

Name

0

1

(unused – reserved)

Unused

1

2

(unused – reserved)

Unused

2

4

Temperature under range ( TEMP)

TEMPUR

3

8

Temperature over-range (TEMP

TEMPOR

4

16

Strain gauge input under-range

ECOMUR

5

32

Strain gauge input over-range

ECOMOR

6

64

Cell under-range (CRAW)

CRAWUR

7

128

Cell over-range (CRAW)

CRAWOR

8

256

System under-range (SRAW)

SYSUR

9

512

System over-range (SRAW)

SYSOR

10

1024

(unused – reserved)

Unused

11

2048

Load Cell Integrity Error (LCINTEG)

LCINTEG

12

4096

Watchdog Reset

WDRST

13

8192

(unused – reserved)

Unused

14

16384

Brown-Out Reset

BRWNOUT

15

32768

Reboot warning (Normal Power up)

REBOOT