品牌
生产厂家厂商性质
上海所在地
Analyzer Technology | Fiber optic dual-beam ClearView db photometer (see also document #1033) |
---|---|
Fiber Connectors | SMA 905 |
Light Source | Tungsten-Halogen, >4000 hours typical |
Photometric Noise | <50 µ AU 450-2100 nm 1 minute rms |
Photometric Drift | <500 µAU rms/ °C |
Response Time | 1 second, minimum. User settable |
Measurement Range | -16 to 30 Saybolt units (pathlength can be optimized to increase sensitivity) |
Measurement Accuracy | Complies with ASTM methods D156 , D6045 |
Power | 24 VDC, 3 A; 72 watts |
Environmental | 0 – 45°C, 0 – 90%, sun and rain sheltered |
Channels | Up to 2 (two) sample channels and optional turbidity monitoring |