SpecEl Ellipsometer System

SpecEl Ellipsometer System

参考价: 面议

具体成交价以合同协议为准
2023-07-04 07:30:11
48
产品属性
关闭
广州标旗光电科技发展股份有限公司

广州标旗光电科技发展股份有限公司

免费会员
收藏

组合推荐相似产品

产品简介

SpecEl Ellipsometer System

详细介绍

The SpecEl-2000-VIS Ellipsometer measures polarized light reflected from the surface of a substrate to determine the thickness and refractive index of the material as a function of wavelength. The SpecEl is controlled via a PC. Measure refractive index, absorbance and thickness with the touch of a button.

All-in-one Accurate System

The SpecEl houses an integrated light source, a spectrometer and two polarizers fixed to 70°. It also includes a PC with a 32-bit Windows operating system. The SpecEl can detect a single layer as thin as 0.1 nm and up to 5 µm thick. In addition, it can provide refractive indices to 0.005°.

The SpecEl is available for Call for Price.

SpecEl Software and "Recipe" Files

In SpecEl Software, you can configure and save experiment method files for one-step analysis. after creating a "recipe," you can select the recipe to execute the experiment.

Specifications

Wavelength range: 380-780 nm (standard) or 450-900 nm (optional)
Optical resolution: 4.0 nm FWHM
Accuracy: 0.1 nm thickness; 0.005% refractive index
Angle of incidence: 70°
Film thickness: 1-5000 nm for single transparent film
Spot size: 2 mm x 4 mm (standard) or 200 µm x 400 µm (optional)
Sampling time: 3-15 seconds (minimum)
Kinetic logging: 3 seconds
Mechanical tolerance (height): +/- 1.5 mm, angle +/- 1.0°
Number of layers: Up to 32 layers
Reference: Not applicable

 