Model | IM400 | IM400W | IK400 | IK400W |
---|
Performance |
ShakingType | Horizontal Pivot Shaking | Horizontal Shuttle Shaking |
Circulation Type | Forced Air Circulation |
OperatingTemp. Range | 5 to 60°C |
Temp.Adjustment Accuracy | ±0.5°C (at 37°C) |
Temp.Distribution Accuracy | ±1.0°C (at 37°C) |
Max.FlaskSetting Qty. | Erlenmeyer Flask 500ml x 9pcs. (6kg) |
ShakingSpeed/min. | Approx. 30 to 200times/min. |
ShakingWidth | 70mm |
Components |
Door | Without Window | With Window | Without Window | With Window |
ShakingPlate | Either Erlenmeyer Flask or Shaking Flask (To bechosenbyUser) |
Interior | Stainless Steel SUS304 |
Refrigerator | Air Cooled Fully Closed Type Compressor130W(CoolantHFC134a) |
Heater | Iron – Chrome Wire 500W |
Fan | Axial Fan Motor 13/14W (50/60Hz) |
ShakingMotor | Full-close Type 600W |
Heat-resistant Material | Styrene Foam |
Temp.Adjuster | PID Control by Micro Computer |
Temp.Setting / Display | Digital Setting by Up Key / Green LEDThreeDigitDisplay |
Temp.Setting Range | 5 to 60°C (Displayable Range : -25 to 65°C) |
Temp.Resolution | 0.1°C |
ShakingSpeed Setting / Display | Analogue Setting by Volume Knob / Green LEDTwoDigitDisplay |
ShakingSpeed Setting | Setting : 30 to 200times/min. |
ShakingSpeed Resolution | 10 times/min. |
OperatingMode | Unchangeable |
Temp.Sensor | Pt Sensor, Pt100Ω |
AdditionalFunction | Defrost Function (10min. Defrost by Defrost Key On) |
SafetyDevice |
SafetyCountermeasure | SelfDiagnosis Functions (Sensor Fault, Heater Fault,TriacTrouble, MinRelay Fault), Electric Leakage Breaker, DoorSafetySwitch |
Standard |
ExternalDimension | W715 x D700 x H936 mm | W715 x D650 x H936 mm |
Power | AC100V, 50/60Hz, 9A |
Weight | Approx. 125kg |
Shelf | Stainless Punching Metal 1pc.ShelfSupport2pcs. |