$item.Name

首页>分析检测>环境检测>颗粒分析仪

参考价:

  • 面议
SOLAIR 3350Rx portable airborne particle counters

具体成交价以合同协议为准

2017-07-08苏州市
型号
SOLAIR 3350Rx
苏州市华宇净化设备有限公司

免费会员 生产厂家

该企业相似产品
尘埃粒子计数器,风量仪,生物安全柜,净化工作台,浮游细菌采样器,温湿度压差测试仪
产品简介
With fast, reliable sampling of one cubic meter of air in only ten minutes, the SOLAIR 3350Rx portable airborne particle counters .
详细信息

Airborne Particle Counters

100 LPM Flow Rate
Saves time in Cleanroom Testing

International GMPs require the sampling of a full cubic meter of air per location in Grade A environments, for certification / validation testing. Traditional 1 CFM particle counters take over 35 minutes at each location to sample a full cubic meter of air. The SOLAIR 3350 / 5350 and SOLAIR 3350Rx / 5350Rx particle counters perform this testing in just 10 minutes (a time savings of 25 minutes per location!). In comparison to 50 LPM units the SOLAIR 3350 / 5350 and SOLAIR 3350Rx / 5350Rx particle counters save 10 minutes per location.

This equates to time and money saved in terms of testing cleanrooms and getting facilities up and running faster.

ISO 14644-1 defines the requirements for the minimum sample volume per location as a function of the class limit for the largest considered particle size. For some cleanrooms this volume can be extremely high, depending upon the largest considered particle size. With the SOLAIR 3350’s minimum detection of 0.3 micron, cleanroom testing of ISO 2 thru ISO 9 cleanrooms can be done in considerably less time.

 


Portable Products > SOLAIR 3350Rx
SOLAIR 3350Rx
 

With fast, reliable sampling of one cubic meter of air in only ten minutes, the SOLAIR 3350Rx portable airborne particle counters offer sensitivities of 0.3 microns, at a flow rate of 100 liters per minute or 3.53 CFM. Housed in a sleek 316L stainless steel enclosure, the SOLAIR 3350Rx is ideal for use in facilities that require instrument wipe-down.

The built-in 5.7-inch (14.47 cm) color touch screen makes the SOLAIR 3350Rx easy to configure and operate. The instruments can store up to 3,000 records of particle count data from up to 6 channels. All data can be quickly and reliably transferred to a computer, printed on the built-in thermal printer or transferred to a USB Flash Drive.

Multiple interfaces - Ethernet, RS-485, USB and USB Flash Drive - make communicating with these instruments extremely flexible. The SOLAIR 3350Rx can be used as portable instruments or integrated into a facility monitoring system.




 

Solair 3350Rx Datasheets
Letter Size - Lo Res: 409 KB
 
A4 Size - Lo Res: 398 KB
 

 
Features Benefits Applications
  • SOLAIR 3350Rx: 0.3-25.0µm Size Range
  • Extreme Life Laser Diode Technology
    >20 Years MTTF Laser Lifespan @ 0.3µm
  • 100 Liters per Minute or 3.53 CFM Flow Rate
  • Closed-loop Controlled Clean Pump Technology
  • Designed to Enable 21 CFR Part 11 Compliance (With Additional LMS Express Software)
  • Built-in Thermal Printer
  • 316L Stainless Steel Construction
  • USB Flash Drive Data Transfer
  • Multiple Communication Interfaces - RS-485, Ethernet, USB and USB Flash Drive
  • 5.7-inch (14.47 cm) Color Touch Screen
  • Designed for ISO 21501-4 Compliance:
    Unit-to-Unit Accuracy & Repeatability Assured
  • 3000-record Data Storage Memory
  • 200 Alphanumeric Location Labels
  • Open Architecture Connectivity with OPC Server
  • Test Wizard for Cleanroom Pass/Fail Reports: FS-209E, ISO-14644-1 or EU GMP Annex 1
  • Reduced Sample Time:
    (1 Cubic Meter in 10 minutes)
  • 2 Year Warranty
  • Cost Effective Price per Point
  • Easy Integration to External Systems
    with OPC Server
  • Easy Data Transfer with USB Flash Drive
  • Cleanroom Monitoring
  • Cleanroom Certification for EU GMP or ISO 14644
  • Facility Monitoring
  • Cleanroom Performance Verification
  • Trend Analysis
  • Routine Cleanroom Testing
  • Cleanroom Packaging
  • Filter Testing
 
 
Specifications
Size Range:
 
0.3 - 25.0μm
Channel Sizes:
 
0.3, 0.5, 1.0, 3.0, 5.0, 10.0μm
0.3, 0.5, 1.0, 5.0, 10.0, 25.0μm
Flow Rate:
 
100 LPM or 3.53 CFM
Counting Efficiency:
 
3350Rx = 50% @ 0.3μm; for particles > 0.45μm, per ISO 21501-4
Light Source: Extreme Life Laser Diode Technology with >20 Year Mean Time To Failure (MTTF)
Zero Count Level:
 
<1 count / 5 minutes (per JIS)
Concentration Limits: 200,000 particles / ft3 (7,060,00 particles / m3) @ 5% coincidence loss. Suitable for all Cleanrooms ISO Classes 1 thru 9
Calibration:
 
Factory Calibrated to ISO 21501-4 including Resolution and Counting Efficiency tests using NIST-traceable particles
Count Modes:
 
Manual, Automatic, Concentration and Real Time (with LMS Pharma Edition or LMS Express 7.0 RT)
Data Storage:
 
3,000 Sample Records (Rotating Buffer) which includes particle data, environmental data, locations and times
Communication Modes:
 
RS-485/MODBUS; RS-232; Ethernet TCP/IP; Universal Serial Bus; USB Flash Drive
Supporting Software:
 
LMS XChange, LMS Express 7.0, LMS Express RT, LMS Pharma Edition, ANY OPC Client (with LMS Express 7.0 OPC Server)
Touch Screen Display:
 
5.7-inch (14.47 cm) color, 320x240
Printer: Built-in Thermal Printer
Reports: ISO-14644-1, FS-209E & EU GMP Annex 1
Alarm: Internal Adjustable Alarm, Alarms on Counts and Low Battery
Enclosure:
 
316L Stainless Steel
Sample Output: Internally filtered to HEPA standards (>99.997% @ 0.3μm)
Vacuum Source:
 
Internal closed-loop flow-controlled pump
Power: 100-240 VAC, 50-60 Hz
Battery: Li-Ion, Removable & Rechargeable: Typical Battery Operation - *4 hours; Typical Re-charge Time - 4 hours
Dimensions:
 
14.77” (d) x 8.0” (w) x 10.41” (h) [37.52 x 20.32 x 26.44 cm]
Weight: With Battery: 25.0 lbs (11.34 kg) Without Battery: 22.0 lbs (9.98 kg)
  * typical battery operation of 10-minute sample, 10-minute hold time, repeat
Environmental Conditions
Operating: 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing
Storage: 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing
Accessories
Included: Operating Manual on CD; Data Cable; Isokinetic Sample Probe with Tripod and Tubing; Purge Filter; Removable/ Rechargeable Li-Ion Battery; Power Cable and Spare Fuse; LMS XChange Software; Printer Paper
Optional: Printed Operating Manual; Cleanroom Cart; Cleanroom Thermal Printer Paper; Spare Removable/Rechargeable Li-Ion Battery; External Battery Charger; USB Flash Drive; LMS Express RT Software; LMS Pharma Edition Software; Validation Manual