$item.Name

首页>电工仪表>电力测试仪>电压测试仪

参考价:

  • 面议

TREK520|523|876|884手持式静电电压测试仪美国*

具体成交价以合同协议为准

2018-05-15苏州市
型号
苏州市艾斯科净化设备有限公司

免费会员 生产厂家

该企业相似产品
静电消除器系列,仪器系列,无尘室用品,净化设备,净化工程,电子工具等
产品简介
TREK520|523|876|884手持式静电电压测试仪美国*
详细信息

美国 TREK 520手持式静电电压测试仪
特点:
TRE-520:0~±2KVDC
TRE-523:0~±20KVDC
精准度±5%, 具备Hold功能
应用: 人体或对象表面静面量测试,主要用于厂内静电产生情况的抽检

Model 520 Series HAND-HELD Electrostatic Voltmeter

The Model 520 Series includes the Model 520 and Model 523.

The TREK Model 520 Hand-Held Electrostatic Voltmeter (ESVM) Series products provide accurate, noncontacting measurements of electrostatic surface voltage for ESD applications in either ionized or non-ionized environments. Industrial applications include measuring charge accumulation in the LCD manufacturing process or measuring static charge in semiconductor production. The Model 520 series ESVMs utilize a new measurement technique that overcomes the disadvantage of the typical hand-held fieldmeter by providing surface voltage measurements which are essentially independent of the sensor probe-to-measured surface spacing.

Model:520-1 ESVM
Over a spacing range of 5 mm to 25 mm, the 520 voltmeter provides better than a 5% voltage measurement accuracy, while the typical fieldmeter provides a voltage measurement error in excess of 50%. An optional voltage monitor output is available (Model:520-2). 

Model 523 ESVM
Over a spacing range of 30 mm to 60 mm, the 523 voltmeter provides greater than a 5% voltage measurement accuracy, while the typical fieldmeter provides a voltage measurement error in excess of 50%. 

The 520 series provides a measurement system which complies with standardized calibration equipment traceable to the National Institute of Standards and Technology. The 520 series hand-held voltmeters employ a slender sensor probe design that provides flexibility to make measurements in many locations which are impossible to reach using the large, bulky sensing surfaces of the typical hand-held fieldmeter. The slender design also allows for higher spatial resolution capability by allowing the measurement of small surface areas. A "hold" pushbutton retains the displayed measured voltage value for recording purposes.

·  Model:520 Measurement Range 
0 to ±2 kV DC.

Model 523 Measurement Range 
0 to ±20 kV DC.

Model:520 Accuracy 
ESVM Monitor (Model:520-2) 
Better than ±5% of full scale over the entire recommended probe-to-surface separation range of 5 mm to 25 mm.

Model 523 Accuracy 
Better than ±5% of full scale over the entire recommended probe-to-surface separation range of 30 mm to 60 mm.

Model:520 Probe-to-Surface Separation (recommended)
5 mm to 25 mm.

Model 523 Probe-to-Surface Separation (recommended)
30 mm to 60 mm.

Stability 
Drift with Time 
Less than 600 ppm/hour, noncumulative.
Drift with Temperature 
Less than 600 ppm/°C.

相关技术文章

同类产品推荐

提示

仪表网采购电话

温馨提示

该企业已关闭在线交流功能