测量原理 | 雷达物位仪 |
特点/应用 | For basic level measurement in liquids, pastes and slurries; non affected by changing media, temperature changes, gas blankets or vapor; Encapsulated PVDF or PP clad horn antenna |
特点 | SIL 2 according to IEC 61508, Safety and reliability with Multi-Echo Tracking, HistoROM |
电源/通信 | 2-wire (HART / PROFIBUS PA/ FOUNDATION Fieldbus) 4-wire (HART) |
频率 | K-band (~26 GHz) |
测量精度 | +/- 2 mm (0.08 in) |
环境温度 | -40...+80 °C (-40...+176 °F) |
过程温度 | -40...+130 °C (-40...+266 °F) |
过程压力(绝压)/过压限定值 | Vacuum...3 bar (Vacuum...43.5 psi) |
主要接液部件 | PVDF, PTFE, Viton, PP, PBT |
过程连接 | Thread: G1 1/2, MNPT1 1/2 Flange: UNI DN80...DN150 (3"...6") |
测量距离 | Standard: 30 m (98 ft) With enhanced dynamics: 40 m (131 ft) |
| |
通信 | 4...20 mA HART PROFIBUS PA FOUNDATION Fieldbus |
证书和认证 | ATEX, FM, CSA, CSA C/US, IEC Ex, TIIS, INMETRO, NEPSI, KC |
安全认证 | Overfill prevention WHG SIL |
选项 | Display Customized parameterization PWIS free |
应用限制 | Maximum measuring range is dependent on the tank form and/or application: Ammoniacal gas phase: FMR54 in stilling well Strong buildup formation: FMR54 with air purge Low DK: FMR51 Only PTFE resistant: FMR52 Custody transfer measurement: FMR5xx |