$item.Name

首页>压力仪表>压力表>大气压力表

参考价:

  • 面议
RPM4 BA100K Reference Pressure Barometer

具体成交价以合同协议为准

2022-11-12上海
型号
该企业相似产品
红外测温仪,炉温跟踪系统,红外成像过程系统
产品简介
Transferstandardlevelmeasurementuncertainty,±0
详细信息
  • Transfer standard level measurement uncertainty, ± 0.01 % of reading (± 0.008 % for dual channel mode)
  • Single channel or dual channel with parallel measurementmode to reduce uncertainty and increase reliability
  • Large, easy to read display
  • RS232 and IEEE-488 interfaces included
  • On-board functions including adjustable average time, high/low, deviation from set point, rate, freeze
  • 17025 accredited calibration included in purchase price.
  • Ideal for absolute pressure AutoZeroing of pressure transfer standards and molbox flow terminals
  • Compact and rugged presentation
  • Optional battery pack

The RPM4 BA100K is a full function laboratory quality barometer intended to measure atmospheric pressure with the very highest performance. Its outstanding pressure measurement specifications are made possible by DHI's exclusive quartz reference pressure transducer (Q-RPT) modules.

  RPM4 BA100K (Single Channel) RPM4 BA100K/BA100K (Dual Channel)
Range 70 kPa to 110 kPa (10 psia to 16 psia)
Power requirements 85 V ac to 264 V ac, 50/60 Hz; 12 V dc, 1.2 A (battery)
Weight 5 kg (11 lb)
Dimensions (H x W x D) 10 cm x 22.7 cm x 24 cm (3.9 in. x 8.9 in. x 9.5 in.)
Test port connections 1/8 in NPT F
Communications ports RS232 (COM1, COM2), IEEE-488.2 Resolution To 1 ppm, user adjustable To 1 ppm, user
Resolution To 1 ppm, user adjustable
Operating temperature 15 °C to 35 °C (59 °F to 95 °F)
Warm up time 30 minute temperature stabilization recommended from cold power up
Vibration Meets MIL-T-28800D
Predicted Stability1 ± 0.005 % of reading ± 0.0032 % of reading
Precision2 ± 0.008 % of reading ± 0.006 % of reading
Measurement Uncertainty3 ± 0.01 % of reading ± 0.008 % of reading

1Predicted Q-RPT measurement stability limit (k=2) over one year assuming regular use of AutoZero function. AutoZero is performed by the operator by comparison with a barometric reference in absol mode. Absolute mode predicted one year stability without AutoZ is: Single Channel ± (5 Pa + 0.005 % of reading), D Channel ± (4 Pa + 0.0032 % of reading).

2Combined linearity, hysteresis, repeatability.

3Maximum deviation of the Q-RPT indication from the true value of applied spressure including precision, predicted one year stability limit, temperature effect and calibration uncertainty, combined and expanded (k=2) following the ISO “Guide to the Expression of Uncertainty in Measurement.”